Biomechanics of the tether breakage: tensile behaviour of a single-unit vertebral body tethering construct.
Guldeniz, Ogulcan; Yip, Christopher C H; Nafo, Wanis; Cheung, Kenneth M C.
Afiliação
  • Guldeniz O; Department of Orthopaedics and Traumatology, The University of Hong Kong, Pokfulam, Hong Kong SAR, China.
  • Yip CCH; Department of Orthopaedics and Traumatology, The University of Hong Kong, Pokfulam, Hong Kong SAR, China.
  • Nafo W; Department of Mechanical System Engineering, Jeonbuk National University, Jeonju, Republic of Korea.
  • Cheung KMC; Department of Orthopaedic Surgery, HKU-Shenzhen Hospital, Shenzhen, China. cheungmc@hku.hk.
Spine Deform ; 11(4): 825-831, 2023 07.
Article em En | MEDLINE | ID: mdl-36763247
ABSTRACT

PURPOSE:

Tether breakage was reported as the most common complication of vertebral body tethering. However, as the literature suggests the physiological loads do not have the potential to cause the failure of the tether. Currently, the biomechanical reason behind the tether breakage is unknown. The current study aims to elucidate the effects of the tension forces on the failure mechanisms of the VBT and provide mechanical justification for how it can be identified radiographically.

METHODS:

Tensile tests (20%/min strain rate) were performed on single-unit VBT samples. Failure modes and mechanical characteristics were reported.

RESULTS:

The failure took place prematurely due to the slippage of the tether at the screw-tether junction where the tether is damaged significantly by the locking cap. Slippage was initiated at 10-13% tensile strain level where the tensile stress and tension force were 50.4 ± 1.5 MPa and 582.2 ± 30.8 N, respectively.

CONCLUSION:

The failure occurs because of high-stress concentrations generated within the locking region which damages the tether surface and leads to the slippage of the tether. We observed that the loads leading to failure are within the physiological limits and may indicate the high likelihood of the tether breakage. The failure mode observed in our study is shown to be the dominant failure mode, and a design improvement on the gripping mechanism is suggested to avoid failure at the screw-tether junction. We observed that the tether elongates 10-13% prior to the breakage, which can be employed as a diagnostic criterion to screen for tether breakages radiographically.