Quantitation of neurofilament light chain protein in serum and cerebrospinal fluid from patients with multiple sclerosis using the MSD R-PLEX NfL assay.
Diagnosis (Berl)
; 10(3): 275-280, 2023 08 01.
Article
em En
| MEDLINE
| ID: mdl-36788117
ABSTRACT
OBJECTIVES:
Neurofilament light (NfL) chain is a marker of neuroaxonal damage in various neurological diseases. Here we quantitated NfL levels in the cerebrospinal fluid (CSF) and serum from patients with multiple sclerosis (MS) and controls, using the R-PLEX NfL assay, which employs advanced Meso Scale Discovery® (MSD) electrochemiluminescence (ECL)-based detection technology.METHODS:
NfL was quantitated in samples from 116 individuals from two sites (Ottawa Hospital Research Institute and Mayo Clinic), consisting of patients with MS (n=71) and age- and sex-matched inflammatory neurological controls (n=13) and non-inflammatory controls (n=32). Correlation of NfL levels between CSF and serum was assessed in paired samples in a subset of MS patients and controls (n=61). Additionally, we assessed the correlation between NfL levels obtained with MSD's R-PLEX® and Quanterix's single molecule array (Simoa®) assays in CSF and serum (n=32).RESULTS:
Using the R-PLEX, NfL was quantitated in 99% of the samples tested, and showed a broad range in the CSF (82-500,000 ng/L) and serum (8.84-2,014 ng/L). Nf-L levels in both biofluids correlated strongly (r=0.81, p<0.0001). Lastly, Nf-L measured by MSD's R-PLEX and Quanterix's Simoa assays were highly correlated for both biofluids (CSF r=0.94, p<0.0001; serum r=0.95, p<0.0001).CONCLUSIONS:
We show that MSD's R-PLEX NfL assay can reliably quantitate levels of NfL in the CSF and serum from patients with MS and controls, where levels correlate strongly with Simoa.Palavras-chave