IMRT QA result prediction via MLC transmission decomposition.
Stasko, John T; Ferris, William S; Adam, David P; Culberson, Wesley S; Frigo, Sean P.
Afiliação
  • Stasko JT; Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Ferris WS; Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Adam DP; Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Culberson WS; Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Frigo SP; Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
J Appl Clin Med Phys ; 24(8): e13990, 2023 Aug.
Article em En | MEDLINE | ID: mdl-37031363
BACKGROUND: Quality assurance measurement of IMRT/VMAT treatment plans is resource intensive, and other more efficient methods to achieve the same confidence are desirable. PURPOSE: We aimed to analyze treatment plans in the context of the treatment planning systems that created them, in order to predict which ones will fail a standard quality assurance measurement. To do so, we sought to create a tool external to the treatment planning system that could analyze a set of MLC positions and provide information that could be used to calculate various evaluation metrics. METHODS: The tool was created in Python to read in DICOM plan files and determine the beam fluence fraction incident on each of seven different zones, each classified based on the RayStation MLC model. The fractions, termed grid point fractions, were validated by analyzing simple test plans. The average grid point fractions, over all control points for 46 plans were then computed. These values were then compared with gamma analysis pass percentages and median dose differences to determine if any significant correlations existed. RESULTS: Significant correlation was found between the grid point fraction metrics and median dose differences, but not with gamma analysis pass percentages. Correlations were positive or negative, suggesting differing model parameter value sensitivities, as well as potential insight into the treatment planning system dose model. CONCLUSIONS: By decomposing MLC control points into different transmission zones, it is possible to create a metric that predicts whether the analyzed plan will pass a quality assurance measurement from a dose calculation accuracy standpoint. The tool and metrics developed in this work have potential applications in comparing clinical beam models or identifying their weak points. Implementing the tool within a treatment planning system would also provide more potential plan optimization parameters.