The impact of software updates on accuracy of intraoral scanners.
BMC Oral Health
; 23(1): 219, 2023 04 15.
Article
em En
| MEDLINE
| ID: mdl-37061664
ABSTRACT
BACKGROUND:
Digital workflow is showing an increasing tendency in everyday dentistry. Accuracy is essential during digital dental workflows for all indication areas. The present study aimed to evaluate the effect of software updates on the accuracy of intraoral scanner (IOS) devices.METHODS:
3Shape Trios 3 Pod with software versions 18.1.2. (TRI3_1) and 20.1.2. (TRI3_2); 3Shape Trios 4 Move, version 19.2.2. (TRI4_1); and 3Shape Trios 4 Pod, version 20.1.1. (TRI4_2) were used to take direct optical impressions from a polymethyl methacrylate (PMMA) full arch reference model with prepared teeth (FDI 11,14,17 for crowns and FDI 26 for onlay) and an edentulous region (between FDI 14 and 17). The scanners were used eight times; STL files were imported into Geomagic Control X for accuracy assessment by comparing them to a reference data set created by an industrial high-precision scanner (AICON SmartScan-3D C5). The average deviation of the surface points was calculated in three locations across a full arch (Parameter 1), the region of a four-unit bridge (Parameter 2), and a single prepared abutment (Parameter 3).RESULTS:
In parameter 1 and 2, the newest model with the latest software (TRI4_2) reached the highest accuracy (31.06 ± 5.24 µm and 21.69 ± 7.50 µm). In parameter 3, an older generation scanner running legacy software produced the highest accuracy TRI4_1, 11.75 ± 0.35 µm.CONCLUSION:
Appropriate software updates can significantly increase the trueness and precision of intraoral scanner devices. With updated software, the older generation can match the accuracy level of latest equipment.Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Técnica de Moldagem Odontológica
/
Desenho Assistido por Computador
Limite:
Humans
Idioma:
En
Ano de publicação:
2023
Tipo de documento:
Article