Your browser doesn't support javascript.
loading
Evaluation of plan complexity and dosimetric plan quality of total marrow and lymphoid irradiation using volumetric modulated arc therapy.
Lambri, Nicola; Dei, Damiano; Hernandez, Victor; Castiglioni, Isabella; Clerici, Elena; De Philippis, Chiara; Loiacono, Daniele; Navarria, Pierina; Reggiori, Giacomo; Rusconi, Roberto; Tomatis, Stefano; Bramanti, Stefania; Scorsetti, Marta; Mancosu, Pietro.
Afiliação
  • Lambri N; Medical Physics Unit, IRCCS Humanitas Research Hospital, Rozzano, Milan, Italy.
  • Dei D; Department of Biomedical Sciences, Humanitas University, Pieve Emanuele, Milan, Italy.
  • Hernandez V; Department of Biomedical Sciences, Humanitas University, Pieve Emanuele, Milan, Italy.
  • Castiglioni I; Radiotherapy and Radiosurgery Department, IRCCS Humanitas Research Hospital, Rozzano, Milan, Italy.
  • Clerici E; Department of Medical Physics, Hospital Universitari Sant Joan de Reus, IISPV, Tarragona, Catalonia, Spain.
  • De Philippis C; Department of Physics "G. Occhialini", University of Milan-Bicocca, Milano, Italy.
  • Loiacono D; Radiotherapy and Radiosurgery Department, IRCCS Humanitas Research Hospital, Rozzano, Milan, Italy.
  • Navarria P; Bone Marrow Transplantation Unit, IRCCS Humanitas Research Hospital, Rozzano, Milan, Italy.
  • Reggiori G; Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Milan, Italy.
  • Rusconi R; Radiotherapy and Radiosurgery Department, IRCCS Humanitas Research Hospital, Rozzano, Milan, Italy.
  • Tomatis S; Medical Physics Unit, IRCCS Humanitas Research Hospital, Rozzano, Milan, Italy.
  • Bramanti S; Department of Biomedical Sciences, Humanitas University, Pieve Emanuele, Milan, Italy.
  • Scorsetti M; Department of Biomedical Sciences, Humanitas University, Pieve Emanuele, Milan, Italy.
  • Mancosu P; IRCCS Humanitas Research Hospital, Rozzano, Milan, Italy.
J Appl Clin Med Phys ; 24(6): e13931, 2023 Jun.
Article em En | MEDLINE | ID: mdl-37085997
ABSTRACT

PURPOSE:

To assess the impact of the planner's experience and optimization algorithm on the plan quality and complexity of total marrow and lymphoid irradiation (TMLI) delivered by means of volumetric modulated arc therapy (VMAT) over 2010-2022 at our institute.

METHODS:

Eighty-two consecutive TMLI plans were considered. Three complexity indices were computed to characterize the plans in terms of leaf gap size, irregularity of beam apertures, and modulation complexity. Dosimetric points of the target volume (D2%) and organs at risk (OAR) (Dmean) were automatically extracted to combine them with plan complexity and obtain a global quality score (GQS). The analysis was stratified based on the different optimization algorithms used over the years, including a knowledge-based (KB) model. Patient-specific quality assurance (QA) using Portal Dosimetry was performed retrospectively, and the gamma agreement index (GAI) was investigated in conjunction with plan complexity.

RESULTS:

Plan complexity significantly reduced over the years (r = -0.50, p < 0.01). Significant differences in plan complexity and plan dosimetric quality among the different algorithms were observed. Moreover, the KB model allowed to achieve significantly better dosimetric results to the OARs. The plan quality remained similar or even improved during the years and when moving to a newer algorithm, with GQS increasing from 0.019 ± 0.002 to 0.025 ± 0.003 (p < 0.01). The significant correlation between GQS and time (r = 0.33, p = 0.01) indicated that the planner's experience was relevant to improve the plan quality of TMLI plans. Significant correlations between the GAI and the complexity metrics (r = -0.71, p < 0.01) were also found.

CONCLUSION:

Both the planner's experience and algorithm version are crucial to achieve an optimal plan quality in TMLI plans. Thus, the impact of the optimization algorithm should be carefully evaluated when a new algorithm is introduced and in system upgrades. Knowledge-based strategies can be useful to increase standardization and improve plan quality of TMLI treatments.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Medula Óssea / Radioterapia de Intensidade Modulada Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Medula Óssea / Radioterapia de Intensidade Modulada Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article