Lesion effects in terms of local impedance variations after pulsed-field ablation during pulmonary vein isolation: a case report.
Iacopino, Saverio; Fabiano, Gennaro; Malacrida, Maurizio; Petretta, Andrea.
Afiliação
  • Iacopino S; Electrophysiology Unit, GVM Care and Research, Maria Cecilia Hospital, Cotignola, RA, Italy.
  • Fabiano G; Electrophysiology Unit, GVM Care and Research, Maria Cecilia Hospital, Cotignola, RA, Italy.
  • Malacrida M; Medical Education & Scientific Affairs, Boston Scientific, Milan, Italy.
  • Petretta A; Electrophysiology Unit, GVM Care and Research, Maria Cecilia Hospital, Cotignola, RA, Italy.
Eur Heart J Case Rep ; 7(5): ytad187, 2023 May.
Article em En | MEDLINE | ID: mdl-37153819
Background: To date, no information is available on highly localized impedance (LI) measurements during the ablation of pulmonary veins (PVs) via a new form of energy such as electroporation by means of pulsed-field ablation (PFA). Case summary: A 55-year-old man with a history of paroxysmal atrial fibrillation was admitted to our hospital for PV isolation (PVI). The procedure was performed with the new multi-electrode PFA catheter (FARAWAVE™). Before energy delivery, a high-density map of the left atrium was constructed with the Rhythmia™ system, while the IntellaNAV Mifi™ OI catheter was used to assess the baseline LI values of the four PVs. A manual tag was used to record the exact position where the IntellaNAV™ catheter measured the LI values for each segment of the vein before and after PVI. The LI values displayed a significant variation after PFA delivery (124.3 ± 5â€…Ω for baseline LI vs. 96.8 ± 6â€…Ω after PFA, P < 0.0001) with a mean absolute LI variation of 27.5 ± 7Ω and a mean percentage LI variation of 25.8 ± 8%. The differences between the average LI values pre- and post-PFA were 28.0 ± 5, 26.5 ± 9, 26.8 ± 3, and 28.8 ± 10â€…Ω for the superior, anterior, posterior, and inferior portions of the PV. Discussion: This is the first instance of the acute characterization, in terms of LI drop, of antral lesions created by a new PFA system. Local impedance variations at ablation sites seem to be larger than those recorded at successful ablation spots obtained by means of thermal energy sources.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article