Dosimetric properties of PASSAG polymer gel dosimeter in electron beam radiotherapy using magnetic resonance imaging.
Zhang, Tiancheng; Almajidi, Yasir Q; Awad, Sameer A; Alhachami, Firas Rahi; Gatea, Maher Abdulfadhil; Kadhum, Wesam R.
Afiliação
  • Zhang T; Department of Radiology, The First People's Hospital of Lianyungang, Lianyungang, China.
  • Almajidi YQ; Baghdad College of Medical Sciences, Baghdad, Iraq.
  • Awad SA; Department of Medical Laboratory Techniques, Al-Maarif University College, Al-Anbar-Ramadi, Iraq.
  • Alhachami FR; Radiology Department, College of Health and Medical Technology, Al-Ayen University, Thi-Qar, Iraq.
  • Gatea MA; Technical Engineering Department College of Technical Engineering, The Islamic University, Najaf, Iraq.
  • Kadhum WR; Department of Pharmacy, Kut University College, Kut, Wasit, Iraq.
J Xray Sci Technol ; 31(4): 825-836, 2023.
Article em En | MEDLINE | ID: mdl-37212060
ABSTRACT

BACKGROUND:

Several physical factors such as photon beam energy, electron beam energy, and dose rate may affect the dosimetric properties of polymer gel dosimeters. The photon beam energy and dose rate dependence of PASSAG gel dosimeter were previously evaluated.

OBJECTIVE:

This study aims to assess the dosimetric properties of the optimized PASSAG gel samples in various electron beam energies.

METHODS:

The optimized PASSAG gel samples are first fabricated and irradiated to various electron energies (5, 7, 10 and 12 MeV). Then, the response (R2) and sensitivity of gel samples are analyzed by magnetic resonance imaging technique at a dose range of 0 to 10 Gy, scanning room temperature range of 15 to 22 °C, and post-irradiation time range of 1 to 30 days.

RESULTS:

The R2-dose response and sensitivity of gel samples do not change under the evaluated electron beam energies (the differences are less than 5%). Furthermore, a dose resolution range of 11 to 38 cGy is obtained for the gel samples irradiated to different electron beam energies. Moreover, the findings show that the R2-dose response and sensitivity dependence of gel samples on electron beam energy varies over different scanning room temperatures and post-irradiation times.

CONCLUSION:

The dosimetric assessment of the optimized PASSAG gel samples provides the promising data for this dosimeter during electron beam radiotherapy.