Serum CRP biomarker detection by using carbon nanotube field-effect transistor (CNT-FET) immunosensor.
Rabbani, Gulam; Khan, Mohammad Ehtisham; Ahmad, Ejaz; Khan, Mohsin Vahid; Ahmad, Abrar; Khan, Anwar Ulla; Ali, Wahid; Zamzami, Mazin A; Bashiri, Abdullateef H; Zakri, Waleed.
Afiliação
  • Rabbani G; IT-medical Fusion Center, 350-27 Gumidae-ro, Gumi-si, Gyeongbuk 39253, Republic of Korea.
  • Khan ME; Department of Chemical Engineering Technology, College of Applied Industrial Technology, Jazan University, 45142, Saudi Arabia. Electronic address: mekhan@jazanu.edu.sa.
  • Ahmad E; Department of Pathology, University of Michigan Medical School, Ann Arbor, MI 48109, USA.
  • Khan MV; Department of Biosciences, Integral University, Lucknow 226026, India.
  • Ahmad A; Department of Biochemistry, Faculty of Science, King Abdulaziz University, Jeddah 21452, Saudi Arabia.
  • Khan AU; Department of Electrical Engineering Technology, College of Applied Industrial Technology, Jazan University, 45142, Saudi Arabia.
  • Ali W; Department of Chemical Engineering Technology, College of Applied Industrial Technology, Jazan University, 45142, Saudi Arabia.
  • Zamzami MA; Department of Biochemistry, Faculty of Science, King Abdulaziz University, Jeddah 21452, Saudi Arabia.
  • Bashiri AH; Department of Mechanical Engineering, College of Engineering, Jazan University, P. O. Box 114, Jazan 45142, Saudi Arabia.
  • Zakri W; Department of Mechanical Engineering, College of Engineering, Jazan University, P. O. Box 114, Jazan 45142, Saudi Arabia.
Bioelectrochemistry ; 153: 108493, 2023 Oct.
Article em En | MEDLINE | ID: mdl-37392576
ABSTRACT
C-reactive protein (CRP) is produced by the liver in response to systemic inflammation caused by bacterial infection, trauma and internal organ failures. CRP serves as a potential biomarker in the precise diagnosis of cardiovascular risk, type-2 diabetes, metabolic syndrome, hypertension and various types of cancers. The pathogenic conditions indicated above are diagnosed by an elevated CRP level in the serum. In this study, we successfully fabricated a highly sensitive and selective carbon nanotube field-effect transistor (CNT-FET) immunosensor for the detection of CRP. The CNTs were deposited on the Si/SiO2 surface, between source-drain electrodes, afterwards modified with well-known linker PBASE and then anti-CRP was immobilized. This anti-CRP functionalized CNT-FET immunosensor exhibits a wide dynamic detection range (0.01-1000 µg/mL) CRP detection, rapid response time (2-3 min) and low variation (<3 %) which can be delivered as a low-cost and rapid clinical detection technology for the early diagnosis of coronary heart disease (CHD). For the clinical applications, our sensor was tested using CRP fortified serum samples and sensing performance was validated using enzyme-linked immune-sorbent assay (ELISA). This CNT-FET immunosensor will be helpful in taking over the complex laboratory-based expensive traditional CRP diagnostic procedures practiced in the hospitals.