Your browser doesn't support javascript.
loading
Predicted vs Observed Valve to Coronary Distance in Valve-in-Valve TAVR: A Computed Tomography Study.
Tzimas, Georgios; Akodad, Mariama; Meier, David; Duchscherer, Jade; Kalk, Kelsey; Everett, Russell J; Haidari, Oliver; Chuang, Ming-Yu A; Sellers, Stephanie L; Dvir, Danny; Sathananthan, Janarthanan; Leipsic, Jonathon A; Webb, John G; Blanke, Philipp.
Afiliação
  • Tzimas G; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada; Service of Cardiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.
  • Akodad M; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
  • Meier D; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
  • Duchscherer J; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
  • Kalk K; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
  • Everett RJ; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
  • Haidari O; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
  • Chuang MA; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
  • Sellers SL; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
  • Dvir D; Shaare Zedek Medical Center, Jerusalem, Israel.
  • Sathananthan J; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
  • Leipsic JA; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
  • Webb JG; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
  • Blanke P; Center for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada. Electronic address: phil.blanke@gmail.com.
JACC Cardiovasc Interv ; 16(16): 2021-2030, 2023 08 28.
Article em En | MEDLINE | ID: mdl-37480892
ABSTRACT

BACKGROUND:

Preprocedural computed tomography (CT) workup with assessment of virtual transcatheter heart valve-to-coronary ostia (VTC) distance and transcatheter heart valve-to-sinus (VTS) distances is recommended to assess the risk of coronary obstruction following valve-in-valve (ViV) transcatheter aortic valve replacement (TAVR).

OBJECTIVES:

The authors sought to investigate the agreement of predicted VTC and VTS distances and observed post-TAVR anatomy on CT and their relationship with transcatheter heart valve (THV) expansion and deployment conditions.

METHODS:

Fifty-one patients who underwent a balloon-expandable ViV procedure were included in this study. The expansion of the THV stent frame was evaluated at 4 levels THV inflow, surgical heart valve (SHV) sewing ring, SHV outflow, and THV outflow. Assessment of the VTC/VTS distances was performed on the pre-TAVR CT, and THV-to-coronary ostia and THV-to-sinus distances were assessed on the post-TAVR CT.

RESULTS:

Following the ViV procedure, the THV stent frame flared toward the outflow but was generally underexpanded at all levels, particularly at the SHV sewing ring level. Postdilatation impacted the extent of THV expansion, resulting in greater expansion than nominal balloon filling at all 4 THV levels (P < 0.001). Observed THV-to-coronary ostia distances were systematically larger than predicted by the VTC distance (mean difference 1.25 ±1.28 mm) in patients with nominal balloon filling but systematically smaller in case of postdilatation (mean difference -0.45 ± 0.52 mm). A similar relationship was observed between VTS and THV-to-sinus distance measurements.

CONCLUSIONS:

With nominal balloon filling, VTC and VTS distances underestimate postprocedural distances due to THV frame underexpansion. However, postdilatation may lead to distances smaller than predicted due to THV overexpansion at the outflow level.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Substituição da Valva Aórtica Transcateter Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Substituição da Valva Aórtica Transcateter Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article