Your browser doesn't support javascript.
loading
Comparison of the Physical Characteristics of Support Stents for Cerebral Aneurysm Embolization.
Iwakami, Takayuki; Fujii, Norio; Son, Jaehyun.
Afiliação
  • Iwakami T; Department of Neurosurgery, Tomishiro Central Hospital, Tomigusuku, Okinawa, Japan.
  • Fujii N; Department of Neurosurgery, Tomishiro Central Hospital, Tomigusuku, Okinawa, Japan.
  • Son J; Department of Neurosurgery, Tomishiro Central Hospital, Tomigusuku, Okinawa, Japan.
J Neuroendovasc Ther ; 15(12): 778-786, 2021.
Article em En | MEDLINE | ID: mdl-37501998
ABSTRACT

Objective:

There is a limited understanding of the characteristics of individual intracranial stents used for aneurysm treatment. We used an experimental model to evaluate the physical characteristics of support stents for aneurysm embolization.

Methods:

Enterprise 2 VRD 4.0 × 39 mm, Neuroform Atlas 4.5 × 21 mm, and LVIS 4.5 × 32 mm stents were 1) observed under light microscopy and subjected to measurements of 2) circumferential radial force, 3) strut tension, 4) stent compression, and 5) conformability upon bending.

Results:

1) Light microscopy showed a large structural difference between laser-cut (Enterprise 2 VRD, Neuroform Atlas) and braided (LVIS) stents. 2) Within the range of indicated blood vessel diameters, the radial force of Enterprise 2 VRD was higher than that of Neuroform Atlas. An extremely large force was required to decrease the LVIS diameter. 3) Neuroform Atlas easily deformed compared to Enterprise 2 VRD, while LVIS was extended with a smaller traction force than that required for Neuroform Atlas. 4) The compression strength was in the order of Enterprise 2 VRD >Neuroform Atlas >LVIS. 5) Enterprise 2 VRD showed a decreased cell area on the concave side, and Neuroform Atlas showed deformation with overlapping struts on the concave side. LVIS naturally adhered to the wall of the blood vessel model.

Conclusion:

Laser-cut and braided stents showed different physical characteristics that were visualized and shown as numerical data. These findings improve the understanding of the proper use of these stents in clinical applications.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Tipo de estudo: Prognostic_studies Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Tipo de estudo: Prognostic_studies Idioma: En Ano de publicação: 2021 Tipo de documento: Article