Fully automated computational measurement of noise in positron emission tomography.
Sartoretti, Thomas; Skawran, Stephan; Gennari, Antonio G; Maurer, Alexander; Euler, André; Treyer, Valerie; Sartoretti, Elisabeth; Waelti, Stephan; Schwyzer, Moritz; von Schulthess, Gustav K; Burger, Irene A; Huellner, Martin W; Messerli, Michael.
Afiliação
  • Sartoretti T; Department of Nuclear Medicine, University Hospital Zurich, Rämistrasse 100, CH-8091, Zurich, Switzerland.
  • Skawran S; University of Zurich, Zurich, Switzerland.
  • Gennari AG; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Zurich, Switzerland.
  • Maurer A; Department of Nuclear Medicine, University Hospital Zurich, Rämistrasse 100, CH-8091, Zurich, Switzerland.
  • Euler A; University of Zurich, Zurich, Switzerland.
  • Treyer V; Department of Nuclear Medicine, University Hospital Zurich, Rämistrasse 100, CH-8091, Zurich, Switzerland.
  • Sartoretti E; University of Zurich, Zurich, Switzerland.
  • Waelti S; Department of Nuclear Medicine, University Hospital Zurich, Rämistrasse 100, CH-8091, Zurich, Switzerland.
  • Schwyzer M; University of Zurich, Zurich, Switzerland.
  • von Schulthess GK; University of Zurich, Zurich, Switzerland.
  • Burger IA; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Zurich, Switzerland.
  • Huellner MW; Department of Nuclear Medicine, University Hospital Zurich, Rämistrasse 100, CH-8091, Zurich, Switzerland.
  • Messerli M; University of Zurich, Zurich, Switzerland.
Eur Radiol ; 34(3): 1716-1723, 2024 Mar.
Article em En | MEDLINE | ID: mdl-37644149
ABSTRACT

OBJECTIVES:

To introduce an automated computational algorithm that estimates the global noise level across the whole imaging volume of PET datasets.

METHODS:

[18F]FDG PET images of 38 patients were reconstructed with simulated decreasing acquisition times (15-120 s) resulting in increasing noise levels, and with block sequential regularized expectation maximization with beta values of 450 and 600 (Q.Clear 450 and 600). One reader performed manual volume-of-interest (VOI) based noise measurements in liver and lung parenchyma and two readers graded subjective image quality as sufficient or insufficient. An automated computational noise measurement algorithm was developed and deployed on the whole imaging volume of each reconstruction, delivering a single value representing the global image noise (Global Noise Index, GNI). Manual noise measurement values and subjective image quality gradings were compared with the GNI.

RESULTS:

Irrespective of the absolute noise values, there was no significant difference between the GNI and manual liver measurements in terms of the distribution of noise values (p = 0.84 for Q.Clear 450, and p = 0.51 for Q.Clear 600). The GNI showed a fair to moderately strong correlation with manual noise measurements in liver parenchyma (r = 0.6 in Q.Clear 450, r = 0.54 in Q.Clear 600, all p < 0.001), and a fair correlation with manual noise measurements in lung parenchyma (r = 0.52 in Q.Clear 450, r = 0.33 in Q.Clear 600, all p < 0.001). Classification performance of the GNI for subjective image quality was AUC 0.898 for Q.Clear 450 and 0.919 for Q.Clear 600.

CONCLUSION:

An algorithm provides an accurate and meaningful estimation of the global noise level encountered in clinical PET imaging datasets. CLINICAL RELEVANCE STATEMENT An automated computational approach that measures the global noise level of PET imaging datasets may facilitate quality standardization and benchmarking of clinical PET imaging within and across institutions. KEY POINTS • Noise is an important quantitative marker that strongly impacts image quality of PET images. • An automated computational noise measurement algorithm provides an accurate and meaningful estimation of the global noise level encountered in clinical PET imaging datasets. • An automated computational approach that measures the global noise level of PET imaging datasets may facilitate quality standardization and benchmarking as well as protocol harmonization.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia por Emissão de Pósitrons Tipo de estudo: Guideline Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia por Emissão de Pósitrons Tipo de estudo: Guideline Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article