Biomechanical Comparison of Different Treatment Strategies for Thoracolumbar Burst Fracture: A Finite Element Study.
Xilong, Cui; Junjun, Zhu; Yuliang, Sun; Wanmei, Yang; Xiumei, Wang; Xiuling, Huang; Haiyang, Yu; Chengmin, Liang; Zikai, Hua.
Afiliação
  • Xilong C; School of Mechatronics Engineering and Automation, Shanghai University, Shanghai, China; Spinal Deformity Clinical and Research Center of Anhui Province, Fuyang, Anhui, China.
  • Junjun Z; School of Mechatronics Engineering and Automation, Shanghai University, Shanghai, China.
  • Yuliang S; School of Mechatronics Engineering and Automation, Shanghai University, Shanghai, China.
  • Wanmei Y; Spinal Deformity Clinical and Research Center of Anhui Province, Fuyang, Anhui, China; Department of Orthopedics, Fuyang People's Hospital, Fuyang, Anhui, China.
  • Xiumei W; School of Mechatronics Engineering and Automation, Shanghai University, Shanghai, China.
  • Xiuling H; School of Mechatronics Engineering and Automation, Shanghai University, Shanghai, China.
  • Haiyang Y; Spinal Deformity Clinical and Research Center of Anhui Province, Fuyang, Anhui, China; Department of Orthopedics, Fuyang People's Hospital, Fuyang, Anhui, China.
  • Chengmin L; Spinal Deformity Clinical and Research Center of Anhui Province, Fuyang, Anhui, China; Department of Orthopedics, Fuyang People's Hospital, Fuyang, Anhui, China.
  • Zikai H; School of Mechatronics Engineering and Automation, Shanghai University, Shanghai, China; Spinal Deformity Clinical and Research Center of Anhui Province, Fuyang, Anhui, China. Electronic address: eddie_hua@shu.edu.cn.
World Neurosurg ; 180: e429-e439, 2023 Dec.
Article em En | MEDLINE | ID: mdl-37757943
ABSTRACT

OBJECTIVE:

The aim of this study was to compare the biomechanical performance of 6 pedicle screw internal fixation strategies for the treatment of burst fractures of the thoracolumbar spine using finite element (FE) analysis.

METHODS:

A finite element model of the T11-L3 thoracolumbar segment was established to simulate L1 vertebral burst fractures, and 6 models were conducted under multidirectional loading conditions P2-D2, P1-D1, P2-D1,P1-D, P1-BF-D1, and P1-UF-D1. The range of motion (ROM) in the T12-L2 region and the von Mises stresses of pedicle screws and rods under the 6 internal fixation models were mainly analyzed.

RESULTS:

The maximum ROM and von Mises stress were obtained under flexion motion in all models. The P1-BF-D1 model had the least ROM and screw stress. However, when the injured vertebra was not nailed bilaterally, the P1-UF-D1 model had the smallest ROM; the maximum von Mises stress on the screw and rod was remarkably higher than that recorded in the other models. Moreover, the P2-D1 model had a ROM similar to that of the P1-D2 model, but with lower screw stress. The 2 models outperformed the P1-D1 model in all 6 conditions. The P2-D2 model had a similar ROM with the P2-D1 model; nevertheless, the maximum von Mises stress was not substantially reduced.

CONCLUSIONS:

The P1-BF-D1 model exhibited better stability and less von Mises stress on the pedicle screws and rods, thereby reducing the risk of screw loosening and fracture. The P2-D1 internal fixation approach is recommended when the fractured vertebrae are not nailed bilaterally.