Efficacy of the positioning system when measuring canine intraocular pressures with the Reichert® Tono-Vera® Vet rebound tonometer.
Cabble, Ava; Bajric, Shayla; Kapeller, Lydia E; Harman, Christine D; Jacobson, Amanda L; Palanivel, Hemalatha; Taylor, David A; Komáromy, András M.
Afiliação
  • Cabble A; Department of Small Animal Clinical Sciences, College of Veterinary Medicine, Michigan State University, East Lansing, Michigan, USA.
  • Bajric S; Department of Small Animal Clinical Sciences, College of Veterinary Medicine, Michigan State University, East Lansing, Michigan, USA.
  • Kapeller LE; Department of Small Animal Clinical Sciences, College of Veterinary Medicine, Michigan State University, East Lansing, Michigan, USA.
  • Harman CD; Department of Small Animal Clinical Sciences, College of Veterinary Medicine, Michigan State University, East Lansing, Michigan, USA.
  • Jacobson AL; Department of Small Animal Clinical Sciences, College of Veterinary Medicine, Michigan State University, East Lansing, Michigan, USA.
  • Palanivel H; Reichert Technologies, Depew, New York, USA.
  • Taylor DA; Reichert Technologies, Depew, New York, USA.
  • Komáromy AM; Department of Small Animal Clinical Sciences, College of Veterinary Medicine, Michigan State University, East Lansing, Michigan, USA.
Vet Ophthalmol ; 27(1): 95-100, 2024 Jan.
Article em En | MEDLINE | ID: mdl-37952123
ABSTRACT

OBJECTIVE:

To compare intraocular pressure (IOP) measurements in dogs taken with the Reichert® Tono-Vera® Vet rebound tonometer with and without the automatic positioning system. ANIMALS STUDIED Measurements were taken on 49 eyes from 26 Beagle-derived dogs with variable genetics-four non-glaucomatous and 22 ADAMTS10-mutant dogs affected with different stages of open-angle glaucoma. Seventeen of the 26 dogs were measured 2-4 times on different days with variable intervals since IOP-lowering medications were administered. PROCEDURES In each dog, tonometry was performed with the Tono-Vera® Vet using three different methods in a randomized order (Method 1) Average of three readings with an automatic positioning system; (Method 2) one reading with an automatic positioning system; and (Method 3) average of three readings obtained without the automatic positioning system. Statistical analyses included one-way ANOVA, Tukey pairwise comparisons, and Bland-Altman plots (MiniTab®).

RESULTS:

With each of the three tonometry methods, 116 measurements were taken, resulting in 348 total IOP measurements with a range of 12.8-49.9 mmHg. The means and standard deviations for each method were 25.4 ± 6.9 mmHg (Method 1), 26.0 ± 7.2 mmHg (Method 2), and 26.9 ± 7.7 mmHg (Method 3), with no significant differences (p = .27). Mean IOP variances were also not significantly different between tonometry methods (p = .24 to .78).

CONCLUSIONS:

Because mean IOPs and their standard deviations were not statistically different between the three tonometry methods, we conclude that Tono-Vera® Vet measurements conducted without the aid of the positioning system still provide reliable results.