Fluid-Dependent Single-Frequency Bioelectrical Impedance Fat Mass Estimates Compared to Digital Imaging and Dual X-ray Absorptiometry.
Nescolarde, Lexa; Orlandi, Carmine; Farina, Gian Luca; Gori, Niccolo'; Lukaski, Henry.
Afiliação
  • Nescolarde L; Department of Electronic Engineering, Universitat Politècnica de Catalunya, 08034 Barcelona, Spain.
  • Orlandi C; Medical Faculty, Tor Vergata University, 00133 Rome, Italy.
  • Farina GL; Medical Center Eubion, 00135 Rome, Italy.
  • Gori N; Medical Center Eubion, 00135 Rome, Italy.
  • Lukaski H; Federazione Italiana Rugby-FIR, Stadio Olimpico, Foro Italico, 00135 Rome, Italy.
Nutrients ; 15(21)2023 Nov 01.
Article em En | MEDLINE | ID: mdl-37960291
The need for a practical method for routine determination of body fat has progressed from body mass index (BMI) to bioelectrical impedance analysis (BIA) and smartphone two-dimensional imaging. We determined agreement in fat mass (FM) estimated with 50 kHz BIA and smartphone single lateral standing digital image (SLSDI) compared to dual X-ray absorptiometry (DXA) in 188 healthy adults (69 females and 119 males). BIA underestimated (p < 0.0001) FM, whereas SLSDI FM estimates were not different from DXA values. Based on limited observations that BIA overestimated fat-free mass (FFM) in obese adults, we tested the hypothesis that expansion of the extracellular water (ECW), expressed as ECW to intracellular water (ECW/ICW), results in underestimation of BIA-dependent FM. Using a general criterion of BMI > 25 kg/m2, 54 male rugby players, compared to 40 male non-rugby players, had greater (p < 0.001) BMI and FFM but less (p < 0.001) FM and ECW/ICW. BIA underestimated (p < 0.001) FM in the non-rugby men, but SLSDI and DXA FM estimates were not different in both groups. This finding is consistent with the expansion of ECW in individuals with excess body fat due to increased adipose tissue mass and its water content. Unlike SLSDI, 50 kHz BIA predictions of FM are affected by an increased ECW/ICW associated with greater adipose tissue. These findings demonstrate the validity, practicality, and convenience of smartphone SLSDI to estimate FM, seemingly not influenced by variable hydration states, for healthcare providers in clinical and field settings.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Composição Corporal / Tecido Adiposo Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Composição Corporal / Tecido Adiposo Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 2023 Tipo de documento: Article