Your browser doesn't support javascript.
loading
A novel sustainable second-derivative synchronous spectrofluorimetric method for the simultaneous determination of olmesartan medoxomil and rosuvastatin calcium in bulk and formulations: Integrated greenness and whiteness evaluation.
Batakoushy, Hany A; Nasr, Zeinab Adel; Soliman, Marwa M; Barghash, Sona S; Fouad, Fatma A; Barseem, Aya.
Afiliação
  • Batakoushy HA; Pharmaceutical Analytical Chemistry Department, Faculty of Pharmacy, Menoufia University, Shebin Elkom, Egypt.
  • Nasr ZA; Pharmaceutical Analytical Chemistry Department, Faculty of Pharmacy (Girls), Al-Azhar University, Cairo, Egypt.
  • Soliman MM; Pharmaceutical Analytical Chemistry Department, Faculty of Pharmacy (Girls), Al-Azhar University, Cairo, Egypt.
  • Barghash SS; Pharmaceutical Analytical Chemistry Department, Faculty of Pharmacy (Girls), Al-Azhar University, Cairo, Egypt.
  • Fouad FA; Pharmaceutical Analytical Chemistry Department, Faculty of Pharmacy (Girls), Al-Azhar University, Cairo, Egypt.
  • Barseem A; Pharmaceutical Analytical Chemistry Department, Faculty of Pharmacy, Menoufia University, Shebin Elkom, Egypt.
Luminescence ; 2023 Dec 07.
Article em En | MEDLINE | ID: mdl-38062646
The present research has established a quick and highly sensitive second-derivative synchronous fluorometric technique for the simultaneous quantification of a binary mixture of olmesartan medoxomil and rosuvastatin calcium. Simultaneously, the suggested approach was used to detect the synchronous fluorescence intensity of the cited drugs at Δ λ = 80 nm in ethanol to determine the concentrations of olmesartan medoxomil and rosuvastatin calcium at 265 and 240 nm, respectively. Various experimental conditions were tested, and each variable was analyzed and optimized. The calibration graphs were shown to be linear within ranges of 0.1-2.0 and 0.5-6.0 µg ml-1 for each drug concentration, respectively. The newly developed Green Solvents Selecting Tool (GSST) was utilized to assess the solvent's sustainability. Furthermore, the proposed method was found to be environmentally friendly after being evaluated with three different tools [the Green Analytical Procedure Index (GAPI), the Analytical Greenness Metric (AGREE), and the Analytical Eco-Scale with Eco-score equal to 95]. The whiteness qualities were also studied using the Red-Green-Blue (RGB12) model, which was recently designed and showed a high score equal to 92.9. The proposed method's good findings, as well as its ongoing sustainability, simplicity, and economy, stimulate its application in QC laboratories.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article