High-temperature ultrafast ChipHPLC-MS.
Weise, Chris; Westphal, Hannes; Warias, Rico; Belder, Detlev.
Afiliação
  • Weise C; Institute of Analytical Chemistry, University Leipzig, Linnéstrasse 3, 04103, Leipzig, Germany.
  • Westphal H; Institute of Analytical Chemistry, University Leipzig, Linnéstrasse 3, 04103, Leipzig, Germany.
  • Warias R; Institute of Analytical Chemistry, University Leipzig, Linnéstrasse 3, 04103, Leipzig, Germany.
  • Belder D; Institute of Analytical Chemistry, University Leipzig, Linnéstrasse 3, 04103, Leipzig, Germany. belder@uni-leipzig.de.
Anal Bioanal Chem ; 416(4): 1023-1031, 2024 Feb.
Article em En | MEDLINE | ID: mdl-38112789
ABSTRACT
Herein, we present a miniaturized chip-based HPLC approach coupled to electrospray ionization mass spectrometry utilizing temperature to achieve high-speed separations. The approach benefits from the low thermal mass of the microfluidic chip and can form an electrospray from the pre-heated mobile phase. With the help of this technology, isothermal and temperature-programmable operations up to 130°C were pursued to perform reversed-phase separations of pesticides in methanol and ethanol-containing eluents in less than 20 s.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article