Dental imaging in clinical photon-counting CT at a quarter of DVT dose.
Sawall, Stefan; Maier, Joscha; Sen, Sinan; Gehrig, Holger; Kim, Ti-Sun; Schlemmer, Heinz-Peter; Schönberg, Stefan O; Kachelrieß, Marc; Rütters, Maurice.
Afiliação
  • Sawall S; Division of X-Ray Imaging and CT, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, Heidelberg 69120, Germany; Medical Faculty, Heidelberg University, Im Neuenheimer Feld 672, Heidelberg 69120, Germany. Electronic address: stefan.sawall@dkfz.de.
  • Maier J; Division of X-Ray Imaging and CT, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, Heidelberg 69120, Germany.
  • Sen S; Department of Orthodontics, University Hospital of Schleswig-Holstein, Arnold-Heller-Straße 3, Kiel 24105, Germany.
  • Gehrig H; Department of Operative Dentistry, University Hospital Heidelberg, Heidelberg University, Im Neuenheimer Feld 400, Heidelberg 69120, Germany.
  • Kim TS; Department of Operative Dentistry, University Hospital Heidelberg, Heidelberg University, Im Neuenheimer Feld 400, Heidelberg 69120, Germany.
  • Schlemmer HP; Division of Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, Heidelberg 69120, Germany.
  • Schönberg SO; Department of Clinical Radiology and Nuclear Medicine, University Hospital Mannheim, Theodor-Kurz-Ufer 1-3, Mannheim 68167, Germany.
  • Kachelrieß M; Division of X-Ray Imaging and CT, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, Heidelberg 69120, Germany; Medical Faculty, Heidelberg University, Im Neuenheimer Feld 672, Heidelberg 69120, Germany.
  • Rütters M; Department of Operative Dentistry, University Hospital Heidelberg, Heidelberg University, Im Neuenheimer Feld 400, Heidelberg 69120, Germany.
J Dent ; 142: 104859, 2024 03.
Article em En | MEDLINE | ID: mdl-38272436
ABSTRACT

OBJECTIVE:

To investigate the image quality of a low-dose dental imaging protocol in the first clinical photon-counting computed tomography (PCCT) system in comparison to a normal-dose acquisition in a digital volume tomography (DVT) system. MATERIALS AND

METHODS:

Clinical PCCT systems offer an increased spatial resolution compared to previous generations of clinical systems. Their spatial resolution is in the order of dental DVT systems. Resolution-matched acquisitions of ten porcine jaws were performed in a PCCT (Naeotom Alpha, Siemens Healthineers) and in a DVT (Orthophos XL, Dentsply Sirona). PCCT images were acquired with 90 kV at a dose of 1 mGy CTDI16 cm. DVT used 85 kV at 4 mGy. Image reconstruction was performed using the standard algorithms of each system to a voxel size of 160 × 160 × 200 µm. The dose-normalized contrast-to-noise ratio (CNRD) was measured between dentine and enamel and dentine and bone. Two readers evaluated overall diagnostic quality of images and quality of relevant structures such as root channels and dentine.

RESULTS:

CNRD is higher in all PCCT acquisitions. CNRD is 37 % higher for the contrast dentine-enamel and 31 % higher for the dentine-bone contrast (p < 0.05). Overall diagnostic image quality was higher for PCCT over DVT (p < 0.02 and p < 0.04 for readers 1 and 2). Quality scores for anatomical structures were higher in PCCT compared to DVT (all p < 0.05). Inter- and intrareader reproducibility were acceptable (all ICC>0.64).

CONCLUSIONS:

PCCT provides an increased image quality over DVT even at a lower dose level and might enable complex dental imaging protocols in the future. CLINICAL

SIGNIFICANCE:

The evolution of photon-counting technology and it's optimization will increasingly move dental imaging towards standardized 3D visualizations providing both minimal radiation exposure and high diagnostic accuracy.