Do Flat Panel Detector C-Arms Decrease Radiation Exposure Compared to Conventional Image Intensifiers?
Gentry, Vance; Farkouh, Ala'a; Chen, Natalie; Amasyali, Akin S; Lee, Jenna; Srikureja, Nathaniel; Keheila, Mohamed; Okhunov, Zhamshid; Baldwin, D Duane.
Afiliação
  • Gentry V; Department of Urology, Loma Linda University Health, Loma Linda, CA.
  • Farkouh A; Department of Urology, Loma Linda University Health, Loma Linda, CA.
  • Chen N; Department of Urology, Loma Linda University Health, Loma Linda, CA.
  • Amasyali AS; Department of Urology, Loma Linda University Health, Loma Linda, CA.
  • Lee J; Department of Urology, Loma Linda University Health, Loma Linda, CA.
  • Srikureja N; Department of Urology, Loma Linda University Health, Loma Linda, CA.
  • Keheila M; Department of Urology, Loma Linda University Health, Loma Linda, CA.
  • Okhunov Z; Department of Urology, Loma Linda University Health, Loma Linda, CA.
  • Baldwin DD; Department of Urology, Loma Linda University Health, Loma Linda, CA. Electronic address: dbaldwin@llu.edu.
Urology ; 185: 150-156, 2024 03.
Article em En | MEDLINE | ID: mdl-38350550
ABSTRACT

OBJECTIVE:

To compare the radiation dose and image quality between flat panel detector (FPD) and traditional image intensifier (II) C-arms at their lowest radiation settings.

METHODS:

In a ureteroscopy simulation using a cadaver model, the radiation exposure was compared between FPD and II at 4 pulses-per-second (pps) using both low dose and automatic exposure control (AEC) settings. Additionally, the lowest dose settings for each machine were compared (4 pps with low dose in the FPD and 1 pps with low dose in the II). Five trials of 5 minutes were conducted for each setting. Four new optically stimulated luminescent dosimeters were used in each trial to record radiation exposure. Ten blinded urologists completed a survey rating image quality for each setting.

RESULTS:

When comparing the FPD and II at their lowest possible settings, the FPD produced significantly more radiation (P <.05). Using both machines at 4 pps in low dose mode resulted in no significant difference between C-arms (P >.05). Conversely, operating the C-arms at 4 pps and AEC resulted in significantly higher radiation exposure from the FPD compared to the II (P <.05). There was no significant difference in image quality at each setting.

CONCLUSION:

FPDs produce significantly more radiation at the lowest settings compared to IIs. Surgeons should employ IIs when reducing radiation exposure as low as possible is imperative, such as when operating on pediatric and pregnant patients.