Effect of skin tone on the accuracy of the estimation of arterial oxygen saturation by pulse oximetry: a systematic review.
Martin, Daniel; Johns, Chris; Sorrell, Lexy; Healy, Eugene; Phull, Mandeep; Olusanya, Segun; Peters, Mark; Fabes, Jeremy.
Afiliação
  • Martin D; Peninsula Medical School, University of Plymouth, John Bull Building, Plymouth, UK; Intensive Care Unit, University Hospitals Plymouth, Plymouth, UK. Electronic address: daniel.martin@plymouth.ac.uk.
  • Johns C; Library & Digital Support, University of Plymouth, Drake Circus, Plymouth, UK.
  • Sorrell L; Peninsula Medical School, University of Plymouth, John Bull Building, Plymouth, UK.
  • Healy E; Dermatopharmacology, Clinical and Experimental Sciences, Faculty of Medicine, University of Southampton, Southampton, UK; Dermatology, University Hospital Southampton NHS Foundation Trust, Southampton, UK.
  • Phull M; Barking, Havering and Redbridge University Trust, Romford, UK; William Harvey Research Institute, Queen Mary University London, London, UK.
  • Olusanya S; St Bartholomew's Hospital, London, UK.
  • Peters M; Paediatric Intensive Care Unit, Great Ormond Street Hospital for Children NHS Foundation Trust and NIHR Biomedical Research Centre, London, UK; University College London Great Ormond St Institute of Child Health, London, UK.
  • Fabes J; Peninsula Medical School, University of Plymouth, John Bull Building, Plymouth, UK; Anaesthetic Department, University Hospitals Plymouth, Plymouth, UK.
Br J Anaesth ; 132(5): 945-956, 2024 May.
Article em En | MEDLINE | ID: mdl-38368234
ABSTRACT

BACKGROUND:

Pulse oximetry-derived oxygen saturation (SpO2) is an estimate of true arterial oxygen saturation (SaO2). The aim of this review was to evaluate available evidence determining the effect of skin tone on the ability of pulse oximeters to accurately estimate SaO2.

METHODS:

Published literature was screened to identify clinical and non-clinical studies enrolling adults and children when SpO2 was compared with a paired co-oximetry SaO2 value. We searched literature databases from their inception to March 20, 2023. Risk of bias (RoB) was assessed using the QUADAS-2 tool. Certainty of assessment was evaluated using the GRADE tool.

RESULTS:

Forty-four studies were selected reporting on at least 222 644 participants (6121 of whom were children) and 733 722 paired SpO2-SaO2 measurements. Methodologies included laboratory studies, prospective clinical, and retrospective clinical studies. A high RoB was detected in 64% of studies and there was considerable heterogeneity in study design, data analysis, and reporting metrics. Only 11 (25%) studies measured skin tone in 2353 (1.1%) participants; the remainder reported participant ethnicity 68 930 (31.0%) participants were of non-White ethnicity or had non-light skin tones. The majority of studies reported overestimation of SaO2 by pulse oximetry in participants with darker skin tones or from ethnicities assumed to have darker skin tones. Several studies reported no inaccuracy related to skin tone. Meta-analysis of the data was not possible.

CONCLUSIONS:

Pulse oximetry can overestimate true SaO2 in people with darker skin tones. The clinical relevance of this bias remains unclear, but its magnitude is likely to be greater when SaO2 is lower. SYSTEMATIC REVIEW PROTOCOL International Prospective Register of Systematic Reviews (PROSPERO) CRD42023390723.