Quantitative measurement of mammographic density in breast-tissue explants using portable NMR: Precision and accuracy.
Foongkajornkiat, Satcha; Sokolowski, Kamil; Stephenson, James; Lloyd, Thomas; Hugo, Honor J; Thompson, Erik W; Momot, Konstantin I.
Afiliação
  • Foongkajornkiat S; School of Chemistry and Physics, Queensland University of Technology, Brisbane, Queensland, Australia.
  • Sokolowski K; Preclincal Imaging Facility, Translational Research Institute, Woolloongabba, Queensland, Australia.
  • Stephenson J; Department of Breast and Endocrine Surgery, Princess Alexandra Hospital, Woolloongabba, Queensland, Australia.
  • Lloyd T; Translational Research Institute, Woolloongabba, Queensland, Australia.
  • Hugo HJ; Department of Diagnostic Radiology, Princess Alexandra Hospital, Woolloongabba, Queensland, Australia.
  • Thompson EW; School of Health and Behavioural Science, University of the Sunshine Coast, Sippy Downs, Queensland, Australia.
  • Momot KI; School of Medicine and Dentistry, Griffith University Sunshine Coast, Birtinya, Queensland, Australia.
Magn Reson Med ; 92(1): 374-388, 2024 Jul.
Article em En | MEDLINE | ID: mdl-38380719
ABSTRACT

PURPOSE:

Single-sided portable NMR (pNMR) has previously been demonstrated to be suitable for quantification of mammographic density (MD) in excised breast tissue samples. Here we investigate the precision and accuracy of pNMR measurements of MD ex vivo as compared with the gold standards.

METHODS:

Forty-five breast-tissue explants from 9 prophylactic mastectomy patients were measured. The relative tissue water content was taken as the MD-equivalent quantity. In each sample, the water content was measured using some combination of three pNMR techniques (apparent T2, diffusion, and T1 measurements) and two gold-standard techniques (computed microtomography [µCT] and hematoxylin and eosin [H&E] histology). Pairwise correlation plots and Bland-Altman analysis were used to quantify the degree of agreement between pNMR techniques and the gold standards.

RESULTS:

Relative water content measured from both apparent T2 relaxation spectra, and diffusion decays exhibited strong correlation with the H&E and µCT results. Bland-Altman analysis yielded average bias values of -0.4, -2.6, 2.6, and 2.8 water percentage points (pp) and 95% confidence intervals of 13.1, 7.5, 11.2, and 11.8 pp for the H&E - T2, µCT - T2, H&E - diffusion, and µCT - diffusion comparison pairs, respectively. T1-based measurements were found to be less reliable, with the Bland-Altman confidence intervals of 27.7 and 33.0 pp when compared with H&E and µCT, respectively.

CONCLUSION:

Apparent T2-based and diffusion-based pNMR measurements enable quantification of MD in breast-tissue explants with the precision of approximately 10 pp and accuracy of approximately 3 pp or better, making pNMR a promising measurement modality for radiation-free quantification of MD.