Your browser doesn't support javascript.
loading
Electrochemical DNA Cleavage Sensing for EcoRV Activity and Inhibition with an ERGO Electrode.
Oh, Da Eun; Kim, Hyun Beom; Kim, Tae Hyun.
Afiliação
  • Oh DE; Department of Chemistry, Soonchunhyang University, Asan 31538, Republic of Korea.
  • Kim HB; Department of Chemistry, Soonchunhyang University, Asan 31538, Republic of Korea.
  • Kim TH; Department of Chemistry, Soonchunhyang University, Asan 31538, Republic of Korea.
Biosensors (Basel) ; 14(2)2024 Jan 30.
Article em En | MEDLINE | ID: mdl-38391992
ABSTRACT
An electrochemically reduced graphene oxide (ERGO) electrode-based electrochemical assay was developed for rapid, sensitive, and straightforward analysis of both activity and inhibition of the endonuclease EcoRV. The procedure uses a DNA substrate designed for EcoRV, featuring a double-stranded DNA (dsDNA) region labeled with methylene blue (MB) and a single-stranded DNA (ssDNA) region immobilized on the ERGO surface. The ERGO electrode, immobilized with the DNA substrate, was subsequently exposed to a sample containing EcoRV. Upon enzymatic hydrolysis, the cleaved dsDNA fragments were detached from the ERGO surface, leading to a decrease in the MB concentration near the electrode. This diminished the electron transfer efficiency for MB reduction, resulting in a decreased reduction current. This assay demonstrates excellent specificity and high sensitivity, with a limit of detection (LOD) of 9.5 × 10-3 U mL-1. Importantly, it can also measure EcoRV activity in the presence of aurintricarboxylic acid, a known inhibitor, highlighting its potential for drug discovery and clinical diagnostic applications.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Clivagem do DNA / Grafite Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Clivagem do DNA / Grafite Idioma: En Ano de publicação: 2024 Tipo de documento: Article