Greffier, Joël; Faby, Sebastian; Pastor, Maxime; Frandon, Julien; Erath, Julien; Beregi, Jean Paul; Dabli, Djamel.
Afiliação
  • Greffier J; IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30029 Nîmes, France. Electronic address: joel.greffier@chu-nimes.fr.
  • Faby S; Department of Computed Tomography, Siemens Healthineers AG, 91301 Forchheim, Germany.
  • Pastor M; IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30029 Nîmes, France.
  • Frandon J; IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30029 Nîmes, France.
  • Erath J; Department of Computed Tomography, Siemens Healthineers AG, 91301 Forchheim, Germany.
  • Beregi JP; IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30029 Nîmes, France.
  • Dabli D; IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30029 Nîmes, France.
Diagn Interv Imaging ; 105(9): 311-318, 2024 Sep.
Article em En | MEDLINE | ID: mdl-38429207
ABSTRACT

PURPOSE:

The purpose of this study was to assess image quality and dose level using a photon-counting CT (PCCT) scanner by comparison with a dual-source CT (DSCT) scanner on virtual monoenergetic images (VMIs) at low energy levels. MATERIALS AND

METHODS:

A phantom was scanned using a DSCT and a PCCT with a volume CT dose index of 11 mGy, and additionally at 6 mGy and 1.8 mGy for PCCT. Noise power spectrum and task-based transfer function were evaluated from 40 to 70 keV on VMIs to assess noise magnitude and noise texture (fav) and spatial resolution on two iodine inserts (f50), respectively. A detectability index (d') was computed to assess the detection of two contrast-enhanced lesions according to the energy level used.

RESULTS:

For all energy levels, noise magnitude values were lower with PCCT than with DSCT at 11 and 6 mGy, but greater at 1.8 mGy. fav values were higher with PCCT than with DSCT at 11 mGy (8.6 ± 1.5 [standard deviation [SD]%), similar at 6 mGy (1.6 ± 1.5 [SD]%) and lower at 1.8 mGy (-17.8 ± 2.2 [SD]%). For both inserts, f50 values were higher with PCCT than DSCT at 11- and 6 mGy for all keV levels, except at 6 mGy and 40 keV. d' values were higher with PCCT than with DSCT at 11- and 6 mGy for all keV and both simulated lesions. Similar d' values to those of the DSCT at 11 mGy, were obtained at 2.25 mGy for iodine insert at 2 mg/mL and at 0.96 mGy for iodine insert at 4 mg/mL at 40 keV.

CONCLUSION:

Compared to DSCT, PCCT reduces noise magnitude and improves noise texture, spatial resolution and detectability on VMIs for all low-keV levels.