Right disc thrombosis of the new Gore Cardioform ASD Occluder.
Catheter Cardiovasc Interv
; 103(5): 741-750, 2024 Apr.
Article
em En
| MEDLINE
| ID: mdl-38451154
ABSTRACT
BACKGROUND:
The GCA is a well performing device in terms of efficacy despite complex anatomies (aortic rim <5 mm and ASD diameter >17 mm) with a good safety profile.AIMS:
To evaluate atrial septal defect (ASD) features impacting on right disc device thrombosis in patients who underwent Gore Cardioform ASD Occluder (GCA) implantation.METHODS:
A total of 44 consecutive patients undergoing percutaneous ASD with GCA device from January 2020 to September 2022 at our tertiary care Center were evaluated. The minimum follow-up was 6 months.RESULTS:
The patients were stratified in two groups according to a cut-off value of ASD diameter equal to 20 mm at sizing balloon, derived from ROC analysis (AUC = 0.894; p = 0.024). Baseline characteristics were comparable between groups in terms of age, sex, weight, height, and interatrial septum dimensions. Patients with ASD > 20 mm (n = 9) had a higher ASD/device dimension ratio, both at echocardiography (p = 0.009) and at sizing balloon (p = 0.001), longer fluoroscopic time (p = 0.022), and higher incidence of device thrombosis (0.006). Right disc thrombosis was observed in three patients of the ASD > 20 mm group, always in the inferior portion of the right disc. On univariate analysis, ASD diameter at sizing balloon (OR 1.360; p = 0.036) was the only positive predictor of device thrombosis.CONCLUSIONS:
Right disc thrombosis of the GCA device may be under-recognized at follow-up, hence deserving clinical attention, especially in those patients with larger ASD diameters.Palavras-chave