Fano cavity test and investigation of the response of the Roos chamber irradiated by proton beams in perpendicular magnetic fields up to 1 T.
Blum, Isabel; Wong, Jing Syuen; Godino Padre, Krishna; Stolzenberg, Jessica; Fuchs, Hermann; Baumann, Kilian-Simon; Poppe, Björn; Looe, Hui Khee.
Afiliação
  • Blum I; University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
  • Wong JS; University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
  • Godino Padre K; University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
  • Stolzenberg J; University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
  • Fuchs H; Division of Medical Physics, Department of Radiation Oncology, Medical University of Vienna, Vienna, Austria.
  • Baumann KS; University Hospital Giessen-Marburg, Department of Radiotherapy and Radiooncology, Marburg, Germany.
  • Poppe B; University of Applied Sciences, Institute of Medical Physics and Radiation Protection, Giessen, Germany.
  • Looe HK; Marburg Ion-Beam Therapy Center, Marburg, Germany.
Phys Med Biol ; 69(8)2024 Apr 03.
Article em En | MEDLINE | ID: mdl-38452383
ABSTRACT
Objective. The aim of this work is to investigate the response of the Roos chamber (type 34001) irradiated by clinical proton beams in magnetic fields.Approach. At first, a Fano test was implemented in Monte Carlo software package GATE version 9.2 (based on Geant4 version 11.0.2) using a cylindrical slab geometry in a magnetic field up to 1 T. In accordance to an experimental setup (Fuchset al2021), the magnetic field correction factorskQB⃗of the Roos chamber were determined at different energies up to 252 MeV and magnetic field strengths up to 1 T, by separately simulating the ratios of chamber signalsMQ/MQB⃗,without and with magnetic field, and the dose-conversion factorsDw,QB⃗/Dw,Qin a small cylinder of water, with and without magnetic field. Additionally, detailed simulations were carried out to understand the observed magnetic field dependence.Main results. The Fano test was passed with deviations smaller than 0.25% between 0 and 1 T. The ratios of the chamber signals show both energy and magnetic field dependence. The maximum deviation of the dose-conversion factors from unity of 0.22% was observed at the lowest investigated proton energy of 97.4 MeV andB⃗= 1 T. The resultingkQB⃗factors increase initially with the applied magnetic field and decrease again after reaching a maximum at around 0.5 T; except for the lowest 97.4 MeV beam that show no observable magnetic field dependence. The deviation from unity of the factors is also larger for higher proton energies, where the maximum lies at 1.0035(5), 1.0054(7) and 1.0069(7) for initial energies ofE0= 152, 223.4 and 252 MeV, respectively.Significance. Detailed Monte Carlo studies showed that the observed effect can be mainly attributed to the differences in the transport of electrons produced both outside and inside of the air cavity in the presence of a magnetic field.