Implantation accuracy of novel polyimide stereotactic electroencephalographic depth electrodes-a human cadaveric study.
Kullmann, Aura; Akberali, Farida; Van Gompel, Jaime J; McGovern, Robert A; Marsh, W Richard; Kridner, Debra; Diaz-Botia, Camilo A; Park, Michael C.
Afiliação
  • Kullmann A; NeuroOne Medical Technologies, Eden Prairie, MN, United States.
  • Akberali F; Zimmer Biomet, Warsaw, IN, United States.
  • Van Gompel JJ; Department of Neurosurgery, Mayo Clinic, Rochester, MN, United States.
  • McGovern RA; Department of Neurosurgery, University of Minnesota Medical Center, Minneapolis, MN, United States.
  • Marsh WR; Department of Neurosurgery, Mayo Clinic, Rochester, MN, United States.
  • Kridner D; NeuroOne Medical Technologies, Eden Prairie, MN, United States.
  • Diaz-Botia CA; NeuroOne Medical Technologies, Eden Prairie, MN, United States.
  • Park MC; Department of Neurosurgery, University of Minnesota Medical Center, Minneapolis, MN, United States.
Front Med Technol ; 6: 1320762, 2024.
Article em En | MEDLINE | ID: mdl-38456122
ABSTRACT

Introduction:

Stereoelectroencephalography (sEEG) is a minimally invasive procedure that uses depth electrodes stereotactically implanted into brain structures to map the origin and propagation of seizures in epileptic patients. Implantation accuracy of sEEG electrodes plays a critical role in the safety and efficacy of the procedure. This study used human cadaver heads, simulating clinical practice, to evaluate (1) neurosurgeon's ability to implant a new thin-film polyimide sEEG electrode according to the instructions for use (IFU), and (2) implantation accuracy.

Methods:

Four neurosurgeons (users) implanted 24 sEEG electrodes into two cadaver heads with the aid of the ROSA robotic system. Usability was evaluated using a questionnaire that assessed completion of all procedure steps per IFU and user errors. For implantation accuracy evaluation, planned electrode trajectories were compared with post-implantation trajectories after fusion of pre- and postoperative computer tomography (CT) images. Implantation accuracy was quantified using the Euclidean distance for entry point error (EPE) and target point error (TPE).

Results:

All sEEG electrodes were successfully placed following the IFU without user errors, and post-implant survey of users showed favorable handling characteristics. The EPE was 1.28 ± 0.86 mm and TPE was 1.61 ± 0.89 mm. Long trajectories (>50 mm) had significantly larger EPEs and TPEs than short trajectories (<50 mm), and no differences were found between orthogonal and oblique trajectories. Accuracies were similar or superior to those reported in the literature when using similar experimental conditions, and in the same range as those reported in patients.

Discussion:

The results demonstrate that newly developed polyimide sEEG electrodes can be implanted as accurately as similar devices in the marker without user errors when following the IFU in a simulated clinical environment. The human cadaver ex-vivo test system provided a realistic test system, owing to the size, anatomy and similarity of tissue composition to that of the live human brain.