Your browser doesn't support javascript.
loading
Effect of Ventilator Settings on Mechanical Power During Simulated Mechanical Ventilation of Patients With ARDS.
El-Khatib, Mohamad F; Zeinelddine, Salah M; HajAli, Thuraya H; Rizk, Marwan; van der Staay, Matthias; Chatburn, Robert L.
Afiliação
  • El-Khatib MF; Drs El-Khatib and Rizk and Ms HajAli are affiliated with the Department of Anesthesiology and Pain Medicine, American University of Beirut, Beirut, Lebanon. Dr Zeinelddine is affiliated with the Department of Internal Medicine, American University of Beirut, Beirut, Lebanon. Mr van der Staay is affi
  • Zeinelddine SM; Drs El-Khatib and Rizk and Ms HajAli are affiliated with the Department of Anesthesiology and Pain Medicine, American University of Beirut, Beirut, Lebanon. Dr Zeinelddine is affiliated with the Department of Internal Medicine, American University of Beirut, Beirut, Lebanon. Mr van der Staay is affi
  • HajAli TH; Drs El-Khatib and Rizk and Ms HajAli are affiliated with the Department of Anesthesiology and Pain Medicine, American University of Beirut, Beirut, Lebanon. Dr Zeinelddine is affiliated with the Department of Internal Medicine, American University of Beirut, Beirut, Lebanon. Mr van der Staay is affi
  • Rizk M; Drs El-Khatib and Rizk and Ms HajAli are affiliated with the Department of Anesthesiology and Pain Medicine, American University of Beirut, Beirut, Lebanon. Dr Zeinelddine is affiliated with the Department of Internal Medicine, American University of Beirut, Beirut, Lebanon. Mr van der Staay is affi
  • van der Staay M; Drs El-Khatib and Rizk and Ms HajAli are affiliated with the Department of Anesthesiology and Pain Medicine, American University of Beirut, Beirut, Lebanon. Dr Zeinelddine is affiliated with the Department of Internal Medicine, American University of Beirut, Beirut, Lebanon. Mr van der Staay is affi
  • Chatburn RL; Drs El-Khatib and Rizk and Ms HajAli are affiliated with the Department of Anesthesiology and Pain Medicine, American University of Beirut, Beirut, Lebanon. Dr Zeinelddine is affiliated with the Department of Internal Medicine, American University of Beirut, Beirut, Lebanon. Mr van der Staay is affi
Respir Care ; 69(4): 449-462, 2024 Mar 27.
Article em En | MEDLINE | ID: mdl-38538014
ABSTRACT

BACKGROUND:

In recent years, mechanical power (MP) has emerged as an important concept that can significantly impact outcomes from mechanical ventilation. Several individual components of ventilatory support such as tidal volume (VT), breathing frequency, and PEEP have been shown to contribute to the extent of MP delivered from a mechanical ventilator to patients in respiratory distress/failure. The aim of this study was to identify which common individual setting of mechanical ventilation is more efficient in maintaining safe and protective levels of MP using different modes of ventilation in simulated subjects with ARDS.

METHODS:

We used an interactive mathematical model of ventilator output during volume control ventilation (VCV) with either constant inspiratory flow (VCV-CF) or descending ramp inspiratory flow, as well as pressure control ventilation (PCV). MP values were determined for simulated subjects with mild, moderate, and severe ARDS; and whenever MP > 17 J/min, VT, breathing frequency, or PEEP was manipulated independently to bring back MP to ≤ 17 J/min. Finally, the optimum VT-breathing frequency combinations for MP = 17 J/min were determined with all 3 modes of ventilation.

RESULTS:

VCV-CF always resulted in the lowest MPs while PCV resulted in highest MPs. Reductions in VT were the most efficient for maintaining safer and protective MP. At targeted MPs of 17 J/min and maximized minute ventilation, the optimum VT-breathing frequency combinations were 250-350 mL for VT and 32-35 breaths/min for breathing frequency in mild ARDS, 200-350 mL for VT and 34-40 breaths/min for breathing frequency in moderate ARDS, and 200-300 mL for VT and 37-45 breaths/min for breathing frequency for severe ARDS.

CONCLUSIONS:

VCV-CF resulted in the lowest MP. VT was the most efficient for maintaining safe and protective MP in a mathematical simulation of subjects with ARDS. In the context of maintaining low and safe MPs, ventilatory strategies with lower-than-normal VT and higher-than-normal breathing frequency will need to be implemented in patients with ARDS.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Respiração Artificial / Síndrome do Desconforto Respiratório Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Respiração Artificial / Síndrome do Desconforto Respiratório Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article