[Identification method of 2-methylisoborneol and geosmin in water by purge and trap-gas chromatography-mass spectrometry].
Cao, Fanghong; Chen, Kun; Chen, Xiaoxia; Yang, Ling; Zhao, Yongxin; Wu, Pinggu.
Afiliação
  • Cao F; Deqing County Center for Disease Control and Prevention, Deqing 313200, China.
  • Chen K; Deqing County Center for Disease Control and Prevention, Deqing 313200, China.
  • Chen X; Deqing County Center for Disease Control and Prevention, Deqing 313200, China.
  • Yang L; Deqing County Center for Disease Control and Prevention, Deqing 313200, China.
  • Zhao Y; Zhejiang Provincial Center for Disease Control and Prevention, Hangzhou 310051, China.
  • Wu P; Zhejiang Provincial Center for Disease Control and Prevention, Hangzhou 310051, China.
Wei Sheng Yan Jiu ; 53(2): 316-331, 2024 Mar.
Article em Zh | MEDLINE | ID: mdl-38604970
ABSTRACT

OBJECTIVE:

To establishe an analysis and identification method for 2-methylisoborneol(2-MIB) and geosmin(GSM) in water using purge and trap-gas chromatography-mass spectrometry.

METHODS:

The samples were enriched and analyzed using a purge and trap system, followed by the separation on a DB-624(30 m×0.25 mm, 1.4 µm) chromatographic column. Quantification was performed using gas chromatography-mass spectrometry with the selected ion monitoring and internal standard calibration.

RESULTS:

The calibration curves for 2-MIB and GSM showed an excellent linearity in the range of 1 to 100 ng/L with R~2 values greater than 0.999. The detection limit and quantification limit for both 2-MIB and GSM were 0.33 ng/L and 1.0 ng/L, respectively. Spike recovery experiments were further carried on the source water and drinking water at three concentration levels. It showed that the average recoveries were from 82.0% to 111.0% for 2-MIB while 84.0% to 110% for GSM. Additionally, the test precision of 2-MIB and GSM ranged from 1.9% to 7.3% and 1.9% to 5.0%(n=6), respectively. The analysis of multiple samples including the local source water, treated water and distribution network water confirmed the existence of 2-MIB and GSM.

CONCLUSION:

Compared to the national standard(GB/T 5750.8-2023), the proposed method enables fully automated sample introduction and analysis without the extra pre-treatment. It provides the advantages of simplicity, good repeatability and high accuracy.