Your browser doesn't support javascript.
loading
Metronome-guided cochlear implantation for slower and smoother insertions of lateral wall electrodes.
Wimmer, W; de Oliveira, J Veloso; Breitsprecher, T M; Hans, S; Van Rompaey, V; Van de Heyning, P; Dazert, S; Weiss, Nora M.
Afiliação
  • Wimmer W; Department of Otorhinolaryngology, Klinikum Rechts Der Isar, TUM School of Medicine and Health, Technical University of Munich, Munich, Germany.
  • de Oliveira JV; Department of Otorhinolaryngology, Klinikum Rechts Der Isar, TUM School of Medicine and Health, Technical University of Munich, Munich, Germany.
  • Breitsprecher TM; Department of Otorhinolaryngology-Head and Neck Surgery, Ruhr-University Bochum, St. Elisabeth-Hospital Bochum, Bochum, Germany.
  • Hans S; MED-EL, Elektromedizinische Geräte, Innsbruck, Austria.
  • Van Rompaey V; Department of Translational Neurosciences, Faculty of Medicine and Health Sciences, University of Antwerp, Antwerp, Belgium.
  • Van de Heyning P; Department of Otorhinolaryngology and Head & Neck Surgery, Antwerp University Hospital, Antwerp, Belgium.
  • Dazert S; Department of Translational Neurosciences, Faculty of Medicine and Health Sciences, University of Antwerp, Antwerp, Belgium.
  • Weiss NM; Department of Otorhinolaryngology and Head & Neck Surgery, Antwerp University Hospital, Antwerp, Belgium.
Article em En | MEDLINE | ID: mdl-38630273
ABSTRACT

INTRODUCTION:

Achieving a slow and smooth electrode array insertion is paramount for preserving structural and functional integrity during cochlear implantation. This controlled study evaluates the efficacy of a metronome-guided insertion technique in enhancing the smoothness and speed of electrode array insertions.

METHODS:

In a prospective cohort study, patients undergoing cochlear implant surgery between 2022 and 2023 with lateral wall electrode arrays were included. Metronome guidance was delivered through an acoustic signal via headphones during electrode array insertion in cochlear implantation and compared to a control group without metronome-guidance.

RESULTS:

In total, 37 cases were evaluated, including 25 conventional insertions and 12 metronome-guided insertions. The results indicate that metronome-guided insertions were significantly slower (- 0.46 mm/s; p < 0.001) without extending the overall procedure time. This can be attributed to fewer paused sections observed in the metronome-guided technique. Moreover, metronome-guided insertions exhibited superior performance in terms of insertion smoothness and a reduced number of re-gripping events.

CONCLUSIONS:

The findings support the recommendation for the systematic application of metronome guidance in the manual insertion of cochlear implant electrode arrays, emphasizing its potential to optimize surgical outcomes.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article