Your browser doesn't support javascript.
loading
Diagnostic performance of a point of care high-sensitivity cardiac troponin I assay and single measurement evaluation to rule out and rule in acute coronary syndrome.
De Iuliis, Vincenzo; Gabriele, Anna Rita; De Santis, Francesca; De Rugeriis, Roberta; Di Quinzio, Luigi; Aloisi, Steeve; Rosati, Anna Carla; Benvenuto, Manuela; Fabiani, Donatello; Chiatamone Ranieri, Sofia.
Afiliação
  • De Iuliis V; Department of Clinical Pathology, 92708 G. Mazzini Civil Hospital , Teramo, Italy.
  • Gabriele AR; Emergency Department, 92708 G. Mazzini Civil Hospital , Teramo, Italy.
  • De Santis F; Department of Clinical Pathology, 92708 G. Mazzini Civil Hospital , Teramo, Italy.
  • De Rugeriis R; Department of Clinical Pathology, 92708 G. Mazzini Civil Hospital , Teramo, Italy.
  • Di Quinzio L; Emergency Department, 92708 G. Mazzini Civil Hospital , Teramo, Italy.
  • Aloisi S; Emergency Department, 92708 G. Mazzini Civil Hospital , Teramo, Italy.
  • Rosati AC; Cardiology Division, 92708 G. Mazzini Civil Hospital , Teramo, Italy.
  • Benvenuto M; Cardiology Division, 92708 G. Mazzini Civil Hospital , Teramo, Italy.
  • Fabiani D; Cardiology Division, 92708 G. Mazzini Civil Hospital , Teramo, Italy.
  • Chiatamone Ranieri S; Department of Clinical Pathology, 92708 G. Mazzini Civil Hospital , Teramo, Italy.
Clin Chem Lab Med ; 2024 Apr 29.
Article em En | MEDLINE | ID: mdl-38669209
ABSTRACT

OBJECTIVES:

About 10 million individuals in USA presented annually in the emergency department (ED) with chest pain or with signs and symptoms of acute coronary syndrome (ACS). The advent of point of care (POC) devices, able to measure high sensitivity troponin, are a very interesting tool in the ED setting for its rapid turnaround time (<10 min).

METHODS:

The present study evaluates the diagnostic performance of the Atellica VTLi (Siemens) in real life setting using the clinical data derived from integrated diagnoses of emergency room staff and cardiologist and in comparison with standard laboratory hs-cTnT assay (Cobas 8000, Elecsys, Roche). 966 patients admitted to the emergency department of "G. Mazzini Hospital" in Teramo, Italy, from July 27, 2022, through June 09, 2023, were enrolled.

RESULTS:

The diagnostic performance of POC hs-cTnI was evaluated. An appropriate POC hs-cTnI threshold values <4 ng/L supplied a sensitivity and an NPV of 100 % (95 % CI 99.5-100) in order to achieve rapid rule out for MI through a single measurement at patient presentation in the ED. Furthermore, a derivation POC hs-cTnI concentration >54 ng/L provided a specificity of 97.2 % (95 % CI 95.9-98.1) and a PPV of 43.5 % (95 % CI 40.3-46.7) for ruling in MI.

CONCLUSIONS:

This platform showed comparable diagnostic performance for myocardial infarction to the central laboratory. Our data suggest the possible use of the Atellica VTLi hs-cTnI POC assay either in emergency department of urban medical centre, either in rural hospital for triage and patient management.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article