Application of stoichiometric CT number calibration method for dose calculation of tissue heterogeneous volumes in boron neutron capture therapy.
Med Phys
; 51(6): 4413-4422, 2024 Jun.
Article
em En
| MEDLINE
| ID: mdl-38669482
ABSTRACT
BACKGROUND:
Monte Carlo simulation code is commonly used for the dose calculation of boron neutron capture therapy. In the past, dose calculation was performed assuming a homogeneous mass density and elemental composition inside the tissue, regardless of the patient's age or sex. Studies have shown that the mass density varies with patient to patient, particularly for those that have undergone surgery or radiotherapy. A method to convert computed tomography numbers into mass density and elemental weights of tissues has been developed and applied in the dose calculation process using Monte Carlo codes. A recent study has shown the variation in the computed tomography number between different scanners for low- and high-density materials.PURPOSE:
The aim of this study is to investigate the effect of the elemental composition inside each calculation voxel on the dose calculation and the application of the stoichiometric CT number calibration method for boron neutron capture therapy planning.METHODS:
Monte Carlo simulation package Particle and Heavy Ion Transport code System was used for the dose calculation. Firstly, a homogeneous cubic phantom with the material set to ICRU soft tissue (four component), muscle, fat, and brain was modelled and the NeuCure BNCT system accelerator-based neutron source was used. The central axis depth dose distribution was simulated and compared between the four materials. Secondly, a treatment plan of the brain and the head and neck region was simulated using a dummy patient dataset. Three models were generated; (1) a model where only the fundamental materials were considered (simple model), a model where each voxel was assigned a mass density and elemental weight using (2) the Nakao20 model, and (3) the Schneider00 model. The irradiation conditions were kept the same between the different models (irradiation time and irradiation field size) and the near maximum (D1%) and mean dose to the organs at risk were calculated and compared.RESULTS:
A maximum percentage difference of approximately 5% was observed between the different materials for the homogeneous phantom. With the dummy patient plan, a large dose difference in the bone (greater than 12%) and region near the low-density material (mucosal membrane, 7%-11%) was found between the different models.CONCLUSIONS:
A stoichiometric CT number calibration method using the newly developed Nakao20 model was applied to BNCT dose calculation. The results indicate the importance of calibrating the CT number to elemental composition for each individual CT scanner for the purpose of BNCT dose calculation along with the consideration of heterogeneity of the material composition inside the defined region of interest.Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Dosagem Radioterapêutica
/
Planejamento da Radioterapia Assistida por Computador
/
Tomografia Computadorizada por Raios X
/
Método de Monte Carlo
/
Terapia por Captura de Nêutron de Boro
/
Imagens de Fantasmas
Limite:
Humans
Idioma:
En
Ano de publicação:
2024
Tipo de documento:
Article