Your browser doesn't support javascript.
loading
Determination of optimum levels of binding antibody units (BAU) of new quantitative chemiluminescent immuno-assay (CLIA) in COVID-19 vaccinated volunteer blood donors.
Tiwari, Aseem Kumar; Aggarwal, Geet; Kale, Pratibha; Yadav, Neeti; Kumar, Virendra; Singh, Gargi; Cheirmaraj, K; Krishnan, K.
Afiliação
  • Tiwari AK; Department of Transfusion Medicine Medanta-The Medicity, Sector-38, Gurgaon, India. Electronic address: draseemtiwari@gmail.com.
  • Aggarwal G; Department of Transfusion Medicine Medanta-The Medicity, Sector-38, Gurgaon, India.
  • Kale P; Department of Clinical Microbiology Institute of Liver and Biliary Sciences, New Delhi, India.
  • Yadav N; Department of Transfusion Medicine Medanta-The Medicity, Sector-38, Gurgaon, India.
  • Kumar V; Department of Microbiology College of Life Sciences, Jiwaji University Gwalior, India.
  • Singh G; Medanta Institute of Education and Research Medanta-The Medicity, Sector-38, Gurgaon, India.
  • Cheirmaraj K; Independent Laboratory Consultant, Chennai, Tamil Nadu, India.
  • Krishnan K; Laboratory Medicine, Thane West, Mumbai 400606, India.
Transfus Apher Sci ; 63(3): 103937, 2024 Jun.
Article em En | MEDLINE | ID: mdl-38678985
ABSTRACT

BACKGROUND:

For assessment of COVID-19 vaccine efficacy, neutralization activity of anti-SARS-CoV-2 antibody is measured. This study was undertaken to determine optimum levels of binding antibody units (BAU/ml) in new quantitative chemiluminescent assay (CLIA) that corresponded to neutralizing potential (30% inhibition) of sVNT assay.

METHODS:

Ninety-one blood samples were analyzed by CLIA and sVNT assays. Test samples (n = 75) were collected from blood donors post-2nd vaccination dose, while control samples (n = 16) were archived pre-COVID donor samples. Correlation between CLIA and sVNT was calculated and receiver operating characteristic (ROC) curve was drawn and analyzed.

RESULTS:

Results indicated excellent correlation between 57.5 BAU/ml on CLIA and 30%inhibition on sVNT assay. ROC curve analysis revealed that the area under the curve (AUC) was 0.971.

DISCUSSION:

The present study determined that 57.5 BAU/ml on CLIA corresponded to 30% inhibition on sVNT assay. Periodic quantitative analysis.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doadores de Sangue / Vacinas contra COVID-19 / SARS-CoV-2 / COVID-19 / Medições Luminescentes / Anticorpos Antivirais Limite: Female / Humans / Male Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doadores de Sangue / Vacinas contra COVID-19 / SARS-CoV-2 / COVID-19 / Medições Luminescentes / Anticorpos Antivirais Limite: Female / Humans / Male Idioma: En Ano de publicação: 2024 Tipo de documento: Article