Feasibility of patient-specific quality assurance (PSQA) for real-time robotic stereotactic body radiotherapy (SBRT) based on tumor motion traces.
J Appl Clin Med Phys
; 25(8): e14352, 2024 Aug.
Article
em En
| MEDLINE
| ID: mdl-38696697
ABSTRACT
PURPOSE:
To design a patient specific quality assurance (PSQA) process for the CyberKnife Synchrony system and quantify its dosimetric accuracy using a motion platform driven by patient tumor traces with rotation.METHODS:
The CyberKnife Synchrony system was evaluated using a motion platform (MODUSQA) and a SRS MapCHECK phantom. The platform was programed to move in the superior-inferior (SI) direction based on tumor traces. The detector array housed by the StereoPhan was placed on the platform. Extra rotational angles in pitch (head down, 4.0° ± 0.15° or 1.2° ± 0.1°) were added to the moving phantom to examine robot capability of angle correction during delivery. A total of 15 Synchrony patients were performed SBRT PSQA on the moving phantom. All the results were benchmarked by the PSQA results based on static phantom.RESULTS:
For smaller pitch angles, the mean gamma passing rates were 99.75% ± 0.87%, 98.63% ± 2.05%, and 93.11% ± 5.52%, for 3%/1 mm, 2%/1 mm, and 1%/1 mm, respectively. Large discrepancy in the passing rates was observed for different pitch angles due to limited angle correction by the robot. For larger pitch angles, the corresponding mean passing rates were dropped to 93.00% ± 10.91%, 88.05% ± 14.93%, and 80.38% ± 17.40%. When comparing with the static phantom, no significant statistic difference was observed for smaller pitch angles (p = 0.1 for 3%/1 mm), whereas a larger statistic difference was observed for larger pitch angles (p < 0.02 for all criteria). All the gamma passing rates were improved, if applying shift and rotation correction.CONCLUSIONS:
The significance of this work is that it is the first study to benchmark PSQA for the CyberKnife Synchrony system using realistically moving phantoms with rotation. With reasonable delivery time, we found it may be feasible to perform PSQA for Synchrony patients with a realistic breathing pattern.Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Garantia da Qualidade dos Cuidados de Saúde
/
Dosagem Radioterapêutica
/
Planejamento da Radioterapia Assistida por Computador
/
Estudos de Viabilidade
/
Radiocirurgia
/
Imagens de Fantasmas
/
Radioterapia de Intensidade Modulada
Limite:
Humans
Idioma:
En
Ano de publicação:
2024
Tipo de documento:
Article