Lateral response artifact correction method using image stitching technique in radiochromic film dosimetry.
Miura, Hideharu; Miyazawa, Masanori; Ozawa, Shuichi; Enosaki, Tsubasa; Kagemoto, Masayuki.
Afiliação
  • Miura H; Hiroshima High-Precision Radiotherapy Cancer Center, Hiroshima, Japan.
  • Miyazawa M; Department of Radiation Oncology, Institute of Biomedical & Health Sciences, Hiroshima University, Hiroshima, Japan.
  • Ozawa S; R-TECH.INC, Tokyo, Japan.
  • Enosaki T; Hiroshima High-Precision Radiotherapy Cancer Center, Hiroshima, Japan.
  • Kagemoto M; Department of Radiation Oncology, Institute of Biomedical & Health Sciences, Hiroshima University, Hiroshima, Japan.
J Appl Clin Med Phys ; 25(8): e14373, 2024 Aug.
Article em En | MEDLINE | ID: mdl-38696704
ABSTRACT

PURPOSE:

Lateral response artifact (LRA) is caused by the interaction between film and flatbed scanner in the direction perpendicular to the scanning direction. This can significantly affect the accuracy of patient-specific quality assurance (QA) in cases involving large irradiation fields. We hypothesized that by utilizing the central area of the flatbed scanner, where the magnitude of LRA is relatively small, the LRA could be mitigated effectively. This study proposes a practical solution using the image-stitching technique to correct LRA for patient-specific QA involving large irradiation fields.

METHODS:

Gafchromic™ EBT4 film and Epson Expression ES-G11000 flatbed scanner were used in this study. The image-stitching algorithm requires a spot between adjacent images to combine them. The film was scanned at three locations on a flatbed scanner, and these images were combined using the image-stitching technique. The combined film dose was then calculated and compared with the treatment planning system (TPS)-calculated dose using gamma analysis (3%/2 mm). Our proposed LRA correction was applied to several films exposed to 18 × 18 cm2 open fields at doses of 200, 400, and 600 cGy, as well as to four clinical Volumetric Modulated Arc Therapy (VMAT) treatment plans involving large fields.

RESULTS:

For doses of 200, 400, and 600 cGy, the gamma analysis values with and without LRA corrections were 95.7% versus 67.8%, 95.5% versus 66.2%, and 91.8% versus 35.9%, respectively. For the clinical VMAT treatment plan, the average pass rate ± standard deviation in gamma analysis was 94.1% ± 0.4% with LRA corrections and 72.5% ± 1.5% without LRA corrections.

CONCLUSIONS:

The effectiveness of our proposed LRA correction using the image-stitching technique was demonstrated to significantly improve the accuracy of patient-specific QA for VMAT treatment plans involving large irradiation fields.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Dosagem Radioterapêutica / Algoritmos / Processamento de Imagem Assistida por Computador / Planejamento da Radioterapia Assistida por Computador / Artefatos / Imagens de Fantasmas / Radioterapia de Intensidade Modulada / Dosimetria Fotográfica Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Dosagem Radioterapêutica / Algoritmos / Processamento de Imagem Assistida por Computador / Planejamento da Radioterapia Assistida por Computador / Artefatos / Imagens de Fantasmas / Radioterapia de Intensidade Modulada / Dosimetria Fotográfica Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article