The significance of partial volume effect on the estimation of hypoxic tumour volume with [18F]FMISO PET/CT.
Kafkaletos, Athanasios; Mix, Michael; Sachpazidis, Ilias; Carles, Montserrat; Rühle, Alexander; Ruf, Juri; Grosu, Anca L; Nicolay, Nils H; Baltas, Dimos.
Afiliação
  • Kafkaletos A; Division of Medical Physics, Department of Radiation Oncology, Medical Center - University of Freiburg, Faculty of Medicine, German Cancer Consortium (DKTK), partner site DKTK-Freiburg, University of Freiburg, Freiburg, Germany. athanasios.kafkaletos@uniklinik-freiburg.de.
  • Mix M; German Cancer Consortium (DKTK), Partner Site Freiburg, a partnership between DKFZ and Medical Center - University of Freiburg, Freiburg, Germany. athanasios.kafkaletos@uniklinik-freiburg.de.
  • Sachpazidis I; Department of Nuclear Medicine, Medical Centre - University of Freiburg, Faculty of Medicine, German Cancer Consortium (DKTK), partner site DKTK-Freiburg, University of Freiburg, Freiburg, Germany.
  • Carles M; Division of Medical Physics, Department of Radiation Oncology, Medical Center - University of Freiburg, Faculty of Medicine, German Cancer Consortium (DKTK), partner site DKTK-Freiburg, University of Freiburg, Freiburg, Germany.
  • Rühle A; Biomedical Imaging Research Group (GIBI230-PREBI) and Imaging La Fe Node at Distributed Network for Biomedical Imaging (ReDIB) Unique Scientific and Technical Infrastructures (ICTS), La Fe Health Research Institute, Valencia, Spain.
  • Ruf J; Department of Radiation Oncology, Medical Center - University of Freiburg, Faculty of Medicine, German Cancer Consortium (DKTK), partner site DKTK-Freiburg, University of Freiburg, Freiburg, Germany.
  • Grosu AL; Department of Radiation Oncology, University of Leipzig Medical Centre, Leipzig, Germany.
  • Nicolay NH; Department of Nuclear Medicine, Medical Centre - University of Freiburg, Faculty of Medicine, German Cancer Consortium (DKTK), partner site DKTK-Freiburg, University of Freiburg, Freiburg, Germany.
  • Baltas D; Department of Radiation Oncology, Medical Center - University of Freiburg, Faculty of Medicine, German Cancer Consortium (DKTK), partner site DKTK-Freiburg, University of Freiburg, Freiburg, Germany.
EJNMMI Phys ; 11(1): 43, 2024 May 09.
Article em En | MEDLINE | ID: mdl-38722446
ABSTRACT

BACKGROUND:

The purpose of this study was to evaluate how a retrospective correction of the partial volume effect (PVE) in [18F]fluoromisonidazole (FMISO) PET imaging, affects the hypoxia discoverability within a gross tumour volume (GTV). This method is based on recovery coefficients (RC) and is tailored for low-contrast tracers such as FMISO. The first stage was the generation of the scanner's RC curves, using spheres with diameters from 10 to 37 mm, and the same homogeneous activity concentration, positioned in lower activity concentration background. Six sphere-to-background contrast ratios were used, from 10.01, down to 2.01, in order to investigate the dependence of RC on both the volume and the contrast ratio. The second stage was to validate the recovery-coefficient correction method in a more complex environment of non-spherical lesions of different volumes and inhomogeneous activity concentration. Finally, we applied the correction method to a clinical dataset derived from a prospective imaging trial (DRKS00003830) forty nine head and neck squamous cell carcinoma (HNSCC) cases who had undergone FMISO PET/CT scanning for the quantification of tumour hypoxia before (W0), 2 weeks (W2) and 5 weeks (W5) after the beginning of radiotherapy. Here, PVE was found to cause an underestimation of the activity in small volumes with high FMISO signal.

RESULTS:

The application of the proposed correction method resulted in a statistically significant increase of both the hypoxic subvolume (171% at W0, 691% at W2 and 4.60 × 103% at W5 with p < 0.001) and the FMISO standardised uptake value (SUV) (27% at W0, 21% at W2 and by 25% at W5 with p < 0.001) within the primary GTV.

CONCLUSIONS:

The proposed PVE-correction method resulted in a statistically significant increase of the hypoxic fraction (HF) with p < 0.001 and demonstrated results in better agreement with published HF data for HNSCC. To summarise, the proposed RC-based correction method can be a useful tool for a retrospective compensation against PVE.