Your browser doesn't support javascript.
loading
Evaluation of output factors of different radiotherapy planning systems using Exradin W2 plastic scintillator detector.
Ando, Yasuharu; Okada, Masahiro; Matsumoto, Natsuko; Ikuhiro, Kawasaki; Ishihara, Soichiro; Kiriu, Hiroshi; Tanabe, Yoshinori.
Afiliação
  • Ando Y; Hiroshima City Hospital, Hiroshima, Japan.
  • Okada M; Hiroshima City North Medical Center Asa Citizens Hospital, Hiroshima, Japan.
  • Matsumoto N; Hiroshima City North Medical Center Asa Citizens Hospital, Hiroshima, Japan.
  • Ikuhiro K; Hiroshima City North Medical Center Asa Citizens Hospital, Hiroshima, Japan.
  • Ishihara S; Hiroshima City Hospital, Hiroshima, Japan.
  • Kiriu H; Hiroshima City Hospital, Hiroshima, Japan.
  • Tanabe Y; Department of Radiological Technology, Graduate School of Health Sciences, Okayama University, 5-1 Shikata-cho, 2-chome, Kita-ku, Okayama, 700-8558, Japan. tanabey@okayama-u.ac.jp.
Phys Eng Sci Med ; 47(3): 1177-1189, 2024 Sep.
Article em En | MEDLINE | ID: mdl-38753285
ABSTRACT
This study aims to evaluate the output factors (OPF) of different radiation therapy planning systems (TPSs) using a plastic scintillator detector (PSD). The validation results for determining a practical field size for clinical use were verified. The implemented validation system was an Exradin W2 PSD. The focus was to validate the OPFs of the small irradiation fields of two modeled radiation TPSs using RayStation version 10.0.1 and Monaco version 5.51.10. The linear accelerator used for irradiation was a TrueBeam with three energies 4, 6, and 10 MV. RayStation calculations showed that when the irradiation field size was reduced from 10 × 10 to 0.5 × 0.5 cm2, the results were within 2.0% of the measured values for all energies. Similarly, the values calculated using Monaco were within approximately 2.0% of the measured values for irradiation field sizes between 10 × 10 and 1.5 × 1.5 cm2 for all beam energies of interest. Thus, PSDs are effective validation tools for OPF calculations in TPS. A TPS modeled with the same source data has different minimum irradiation field sizes that can be calculated. These findings could aid in verification of equipment accuracy for treatment planning requiring highly accurate dose calculations and for third-party evaluation of OPF calculations for TPS.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Plásticos / Contagem de Cintilação / Planejamento da Radioterapia Assistida por Computador Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Plásticos / Contagem de Cintilação / Planejamento da Radioterapia Assistida por Computador Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article