Recommendations for using analytical anisotropic algorithm and AcurosXB for epidermal dose calculations in breast radiotherapy from an in vivo Gafchromic film study.
Malhotra, Aria; Carpentier, Emilie E; Duzenli, Cheryl.
Afiliação
  • Malhotra A; BC Cancer, Vancouver, British Columbia, Canada.
  • Carpentier EE; Department of Physics and Astronomy, University of British Columbia, Vancouver, British Columbia, Canada.
  • Duzenli C; BC Cancer, Vancouver, British Columbia, Canada.
J Appl Clin Med Phys ; 25(8): e14416, 2024 Aug.
Article em En | MEDLINE | ID: mdl-38812120
ABSTRACT
BACKGROUND AND

PURPOSE:

This study recommends clinical epidermal dose calculation methods based on in-vivo film measurements and registered skin dose distributions with the Eclipse (Varian Medical Systems) treatment planning system's Analytical Anisotropic Algorithm (AAA) and Acuros XB (AXB) dose calculation algorithms. MATERIALS AND

METHODS:

Eighteen AAA V13.6 breast plans were recalculated using AXB (dose to medium) V13.5 with the same beam parameters and monitor units as in the original plans. These are compared against in-vivo Gafchromic film measurements from the lateral and inferior breast regions. Three skin structures in the treatment planning system are evaluated a surface layer of voxels of the body contour, a 0.2 cm internal skin rind, and a 0.5 cm internal skin rind.

RESULTS:

Systematic shifts are demonstrated between the film measurements of skin dose and the Eclipse dose calculations. On average, the dose to the surface layer of pixels is underestimated by AAA by 8% and overestimated by AXB by 3%. A 5 mm skin rind extended into the body can increase epidermal dose calculations on average by 8% for AAA and 4% for AXB.

CONCLUSION:

This is the first study to register in-vivo skin dose distributions in the breast to the treatment planning system for comparison. Based on the results from this study it is recommended that epidermal dose is calculated with a 0.5 cm skin rind for the AAA algorithm and with rind thickness up to 0.2 cm for the AXB algorithm.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Neoplasias da Mama / Radioterapia de Intensidade Modulada / Dosimetria Fotográfica Limite: Female / Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Neoplasias da Mama / Radioterapia de Intensidade Modulada / Dosimetria Fotográfica Limite: Female / Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article