CT angiography of the lower limbs: combining small field of view with large matrix size and iterative reconstruction further improves image quality of below-the-knee arteries.
Scheschenja, Michael; König, Alexander; Viniol, Simon; Bastian, Moritz B; Wessendorf, Joel; Mahnken, Andreas H.
Afiliação
  • Scheschenja M; Department of Diagnostic & Interventional Radiology, Philipps-University Marburg, Marburg, Germany.
  • König A; Department of Diagnostic & Interventional Radiology, Philipps-University Marburg, Marburg, Germany.
  • Viniol S; Department of Diagnostic & Interventional Radiology, Philipps-University Marburg, Marburg, Germany.
  • Bastian MB; Department of Diagnostic & Interventional Radiology, Philipps-University Marburg, Marburg, Germany.
  • Wessendorf J; Department of Diagnostic & Interventional Radiology, Philipps-University Marburg, Marburg, Germany.
  • Mahnken AH; Department of Diagnostic & Interventional Radiology, Philipps-University Marburg, Marburg, Germany.
Acta Radiol ; 65(7): 774-783, 2024 Jul.
Article em En | MEDLINE | ID: mdl-38841768
ABSTRACT

BACKGROUND:

Image quality and diagnostic accuracy in computed tomography angiography (CTA) reach their limits in imaging of below-the-knee vessels.

PURPOSE:

To evaluate whether image quality in CTA of lower limbs is further improvable by combining side-separate reconstruction with a larger matrix size and whether resulting noise can be compromised with iterative reconstruction (IR). MATERIAL AND

METHODS:

From CTA of the lower extremities of 26 patients (5 women, 21 men; mean age = 68.5 ± 10.3 years), the lower legs were reconstructed side-separately with different reconstruction algorithms and matrix sizes including filtered back projection (FBP) with a 512 × 512 matrix, FBP with a 1024 × 1024 matrix, IR (SAFIRE) with a 512 × 512 matrix, and IR (SAFIRE) with a 1024 × 1024 matrix. A total of 208 CT series were evaluated. Subjective image quality was assessed by two readers using a 5-point Likert scale. Image noise was assessed by measuring signal-to-noise and contrast-to-noise ratios.

RESULTS:

Subjective image quality was rated significantly higher when using a 1024 × 1024 matrix (P < 0.001) and could further be increased with IR. Vessel sharpness was rated significantly better with a larger matrix (P < 0.001). Visible and measured image noise was significantly higher with a 1024 × 1024 matrix but could be reduced by using IR (P < 0.001), even to a level below FBP with a 512 × 512 matrix while reconstructing with a larger matrix (P < 0.001).

CONCLUSION:

Image quality, image noise, and vessel sharpness can be further improved in CTA of the lower extremities with side-separate reconstruction using a 1024 × 1024 matrix size and IR.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Radiográfica Assistida por Computador / Extremidade Inferior / Angiografia por Tomografia Computadorizada Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Radiográfica Assistida por Computador / Extremidade Inferior / Angiografia por Tomografia Computadorizada Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2024 Tipo de documento: Article