Clinical Applications for Spread of Excitation Functions Obtained Via Electrically Evoked Compound Action Potentials (eCAP).
Berg, Katelyn A; DeFreese, Andrea J; Sisler-Dinwiddie, Allyson L; Labadie, Robert F; Tawfik, Kareem O; Gifford, René H.
Afiliação
  • Berg KA; Department of Hearing and Speech Sciences, Vanderbilt University Medical Center, Nashville, Tennessee.
  • DeFreese AJ; Department of Hearing and Speech Sciences, Vanderbilt University Medical Center, Nashville, Tennessee.
  • Sisler-Dinwiddie AL; Department of Hearing and Speech Sciences, Vanderbilt University Medical Center, Nashville, Tennessee.
  • Labadie RF; Department of Otolaryngology-Head and Neck Surgery, Medical University of South Carolina, Charleston, South Carolina.
  • Gifford RH; Department of Hearing and Speech Sciences, Vanderbilt University Medical Center, Nashville, Tennessee.
Otol Neurotol ; 2024 Jun 26.
Article em En | MEDLINE | ID: mdl-38923968
ABSTRACT

OBJECTIVES:

To assess the clinical utility of spread of excitation (SOE) functions obtained via electrically evoked compound action potentials (eCAP) to 1) identify electrode array tip fold-over, 2) predict electrode placement factors confirmed via postoperative computed tomography (CT) imaging, and 3) predict postoperative speech recognition through the first year post-activation in a large clinical sample. STUDY

DESIGN:

Retrospective case review.

SETTING:

Cochlear implant (CI) program at a tertiary medical center. PATIENTS Two hundred seventy-two ears (238 patients) with Cochlear Ltd. CIs (mean age = 46 yr, range = 9 mo-93 yr, 50% female) implanted between August 2014 and December 2022 were included. MAIN OUTCOME

MEASURES:

eCAP SOE widths (mm) (probe electrodes 5, 11, and 17), incidence of electrode tip fold-over, CT imaging data (electrode-to-modiolus distance, angular insertion depth, scalar location), and speech recognition outcomes (consonant-nucleus-consonant [CNC], AzBio quiet, and +5 dB SNR) through the first year after CI activation.

RESULTS:

1) eCAP SOE demonstrated a sensitivity of 85.7% for identifying tip fold-over instances that were confirmed by CT imaging. In the current dataset, the tip fold-over incidence rate was 3.1% (7 patients), with all instances involving a precurved electrode array. 2) There was a significant positive relationship between eCAP SOE and mean electrode-to-modiolus distance for precurved arrays, and a significant positive relationship between eCAP SOE and angular insertion depth for straight arrays. No relationships between eCAP SOE and scalar location or cochlea diameter were found in this sample. 3) There were no significant relationships between eCAP SOE and speech recognition outcomes for any measure or time point, except for a weak negative correlation between average eCAP SOE widths and CNC word scores at 6 months post-activation for precurved arrays.

CONCLUSIONS:

In the absence of intraoperative CT or fluoroscopic imaging, eCAP SOE is a reasonable alternative method for identifying electrode array tip fold-over and should be routinely measured intraoperatively, especially for precurved electrode arrays with a sheath.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article