Primary stability of multi-hole acetabular cup combined with posterior column plating for the fixation of complex acetabular fractures in elderly patients: a biomechanical analysis.
Yan, Shuang G; Wang, Zhi; Cui, Yiliang; Hua, Xingyi; Liu, Fanxiao; Li, Di.
Afiliação
  • Yan SG; Department of Orthopaedic Surgery, The First Affiliated Hospital of Anhui Medical University, Hefei, China. yanshuanggen686@hotmail.com.
  • Wang Z; Department of Orthopaedic Surgery, Beijing Oriental Electronics Hospital, Hefei, China.
  • Cui Y; Department of Orthopaedic Surgery, The First Affiliated Hospital of Anhui Medical University, Hefei, China.
  • Hua X; Department of Orthopaedic Surgery, The First Affiliated Hospital of Anhui Medical University, Hefei, China.
  • Liu F; Department of Orthopaedics, Shandong Provincial Hospital, Shandong First Medical University, No. 324, Road Jing Wu Wei Qi, Jinan, 250021, Shandong, China.
  • Li D; Department of Orthopaedic Surgery, The First Affiliated Hospital of Naval Medical University, Changhai Hospital, Shanghai, China. orthlee@163.com.
Article em En | MEDLINE | ID: mdl-38926197
ABSTRACT

BACKGROUND:

Acute total hip arthroplasty has gained increasing popularity in the treatment of complex acetabular fractures in elderly patients, but the biomechanical evidence is lacking. Therefore, we measured the primary stability to evaluate whether acute total hip arthroplasty using a multi-hole acetabular cup and posterior column plating through a single K-L approach can safely be used for treating complexed acetabular fractures in the elderly.

METHODS:

In 18 composite osteoporotic hemipelves, T-type acetabular fractures were treated in three ways CSP (Cup with acetabular screws augmentation combined with posterior plate) group, CP (Cup without acetabular screw combined with posterior plate) group and CSPA (Cup with acetabular screws combined with posterior plate and anterior plate) group. Each specimen was dynamically loaded (300 to 1700 N, 1 Hz). Primary stability of the acetabular cup and fracture gap was evaluated by 3-dimensional (3D) micromotions.

RESULTS:

No significant differences in the 3D-micromotions were observed among the CSP, CP and CSPA groups except the test point on the anterior column of the acetabulum in the CP scenario (CSP 49.33 ± 21.08 µm vs. CP 224.83 ± 52.29 µm, p < 0.001; CSPA 45.50 ± 12.16 µm vs. CP 224.83 ± 52.29 µm, p < 0.001). No significant differences in the fracture gap displacement on the posterior column of acetabulum were observed among the CSP, CP and CSPA groups.

CONCLUSION:

Our results show, that acute total hip arthroplasty using a multi-hole cup with acetabular screws and posterior column plating through single a K-L approach offers good primary stability to allow good osseous integration for treating complex acetabular fractures in the elderly. Furthermore, it also provides good fracture gap displacement on the posterior column.