Your browser doesn't support javascript.
loading
Predictive Accuracy Analysis of a Novel Robotic-Assisted System for Total Knee Arthroplasty:A Prospective Observational Study.
Miao, Haichuan; Zhu, Zhiyong; Wang, Huisheng; Bai, Xizhuang; Li, Xi.
Afiliação
  • Miao H; Dalian Medical University, Dalian, 116044, People's Republic of China.
  • Zhu Z; Department of Sports Medicine and Joint Surgery, The People's Hospital of Liaoning Province, The People's Hospital of China Medical University, Shenyang, 110000, People's Republic of China.
  • Wang H; Department of Sports Medicine and Joint Surgery, The People's Hospital of Liaoning Province, The People's Hospital of China Medical University, Shenyang, 110000, People's Republic of China.
  • Bai X; Department of Sports Medicine and Joint Surgery, The People's Hospital of Liaoning Province, The People's Hospital of China Medical University, Shenyang, 110000, People's Republic of China.
  • Li X; Department of Sports Medicine and Joint Surgery, The People's Hospital of Liaoning Province, The People's Hospital of China Medical University, Shenyang, 110000, People's Republic of China.
Ther Clin Risk Manag ; 20: 473-482, 2024.
Article em En | MEDLINE | ID: mdl-39135983
ABSTRACT

Background:

Robotic-assisted total knee arthroplasty (RATKA) has been reported to enhance operative decision-making. The purpose of this study was intended to assess the predictive accuracy of bone cuts, lower limb alignment, and component size of a novel system for RATKA preoperatively and intraoperatively.

Methods:

Preoperatively planned bone cuts, limb alignment, and component size were projected using a reconstructed 3D model. Intraoperative bone cuts and postoperative limb alignment were measured. Errors between planned and real bone cuts, limb alignment, and component size were compared.

Results:

The mean absolute errors for bone cuts and alignment were within 1.40mm/1.30° with a standard deviation (SD) of 0.96mm/1.12°. For all errors of bone cuts and alignment compared with the plan, there were no statistically significant differences except for the lateral distal of femoral cuts (P=0.004). The accuracy for predicting the femoral, tibial, and polyethylene component sizes was 100% (48/48), 90% (43/48), and 88% (42/48), respectively. Regarding all mean absolute errors of bone cuts and alignments, no significant differences were observed among surgeons.

Conclusion:

The novel robotically-assisted system for RATKA donated reliable operative decision-making based on the predictive accuracy regardless of the surgeon's level of experience.