Your browser doesn't support javascript.
loading
A potential usefulness of ultra-high-resolution computed tomography in quality assurance of remote after-loading system for cervical cancer.
Kinjyo, Masashi; Nishie, Akihiro; Kudaka, Ryo; Nakano, Shota; Ariga, Takuro.
Afiliação
  • Kinjyo M; Department of Radiology, Graduate School of Medical Science, University of the Ryukyus, 207 Uehara, Nishihara-cho, Nakagami-gun, Okinawa 903-0215, Japan.
  • Nishie A; Department of Radiology, Graduate School of Medical Science, University of the Ryukyus, 207 Uehara, Nishihara-cho, Nakagami-gun, Okinawa 903-0215, Japan.
  • Kudaka R; Division of Radiology, University of the Ryukyus Hospital, 207 Uehara, Nishihara-cho, Nakagami-gun, Okinawa 903-0215, Japan.
  • Nakano S; Canon Medical Systems Corporation, 70-1 Yanagi-cho, Saiwai-ku, Kawasaki, Kanagawa 212-0015, Japan.
  • Ariga T; Department of Radiology, Graduate School of Medical Science, University of the Ryukyus, 207 Uehara, Nishihara-cho, Nakagami-gun, Okinawa 903-0215, Japan.
J Radiat Res ; 65(5): 689-692, 2024 Sep 24.
Article em En | MEDLINE | ID: mdl-39154375
ABSTRACT
Intracavitary brachytherapy with a remote after-loading system (RALS) is performed as a part of radical radiation therapy in cervical cancer. The radiation source is delivered directly through an applicator placed inside the uterus or vagina. Thorough quality control is important to prevent accidents that can lead to serious irradiation error, and an applicator check is one such quality control measure. We experienced a clinical situation in which a small volume of water was observed in the lumen of a post-sterilized applicator on treatment-planning CT. Although the submersion test was negative and no air bubbles emerged from the applicator, ultra-high-resolution computed tomography (U-HRCT) showed a linear crack reaching the inside of the applicator. This abnormality was not identified on treatment-planning CT, which has lower spatial resolution than U-HRCT. In addition, no linear cracks were seen on U-HRCT images of eight other applicators considered to be free from damage. U-HRCT may have superior potential to detect applicator damage and could be useful for quality assurance of the RALS procedure.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Braquiterapia / Tomografia Computadorizada por Raios X / Neoplasias do Colo do Útero Limite: Female / Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Braquiterapia / Tomografia Computadorizada por Raios X / Neoplasias do Colo do Útero Limite: Female / Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article