Your browser doesn't support javascript.
loading
Surface replacement of the hip with the Tharies system. Two to five-year results.
J Bone Joint Surg Am ; 63(7): 1069-77, 1981 Sep.
Article em En | MEDLINE | ID: mdl-7276044
ABSTRACT
The two to five-year clinical results of 200 Tharies surface replacements were comparable to those of conventional total hip replacement. The incidence of systemic complications, dislocations, and sepsis was low. There were ten cases of aseptic and one case of septic loosening. Nine minor secondary procedures were required in the series. In two patients the femoral component shifted asymptomatically into varus angulation, but no revision was needed. Fifty-five per cent of patients for whom serial radiographs were available had some progression of radiolucency, but there was poor correlation between the width of the zone and loosening until the patient became symptomatic. Improved techniques of preparation of the interface and of delivery, compression, and containment of the cement have improved the postoperative radiographic appearance of the more recently treated patients.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Articulação do Quadril / Prótese de Quadril / Artropatias Tipo de estudo: Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 1981 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Articulação do Quadril / Prótese de Quadril / Artropatias Tipo de estudo: Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 1981 Tipo de documento: Article