Your browser doesn't support javascript.
loading
Principles of acetabular fixation in THR with special reference to the "press-fit cup".
Morscher, E.
Afiliação
  • Morscher E; Felix Platter-Hospital, Basel, Switzerland.
Acta Orthop Belg ; 59 Suppl 1: 260-6, 1993.
Article em En | MEDLINE | ID: mdl-8116408
The "press-fit cup" has been developed by the author according to the current well known principles of cementless fixation of acetabular cups in THR. Its concept is to achieve primary stability through press-fit, by flattening the dome of the hemispheric cup and by using a cup size 1.5 mm larger than the reamer used previously. The coating (Sulmesh) consists of a net shell of four layers of orderly oriented wire meshes with specific pore size and porosity volume. The pure titanium wires are bonded together. The histologies of retrieved cups shows extensive bony ingrowth both in animal experiments and in autopsy specimens. In a radiological follow-up of 100 consecutive primary total hip replacements, the press-fit cup showed complete radiological integration in all 3 zones in 97% of cases. Nonprogressive radiolucencies were found in zones I and III in only 3% of cases.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Prótese de Quadril / Acetábulo Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Animals / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 1993 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Prótese de Quadril / Acetábulo Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Animals / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 1993 Tipo de documento: Article