Dosimetry of effective wedge fields produced by an internal wedge.
Med Phys
; 21(1): 69-75, 1994 Jan.
Article
em En
| MEDLINE
| ID: mdl-8164591
ABSTRACT
A procedure is described to calculate the monitor unit ratios required to produce effective wedge fields having a desired wedge angle by combining an internal 60 degrees wedge with an open field. Complementary procedures are derived and demonstrated for calculating the effective wedge dose distributions with wedge angles of 15 degrees, 30 degrees, and 45 degrees using the central axis depth dose data and off-axis ratios of the open field and the 60 degrees wedged field. Measurements at five points on and off the central axis within each field and measurements of the effective wedge factor demonstrated that the calculated wedge distributions were correctly delivered to within 2% in all cases.