Evaluation of truncated readout-echo MR images obtained at 0.35 T.
Avram, H E; Hale, J D; Li, A J; Kaufman, L.
Afiliação
  • Avram HE; Toshiba America MRI, Inc, South San Francisco, Calif.
Radiology ; 186(3): 896-901, 1993 Mar.
Article em En | MEDLINE | ID: mdl-8430206
ABSTRACT
It is possible to acquire a truncated echo, in which part of the information at the beginning or end of the echo is missing, and to restore the missing information by conjugation. This process was shown by means of edge spread function in phantoms and brain images in volunteers to introduce little, if any, degradation of image quality and can be used to improve sequence efficiency or to shorten echo time.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Encéfalo / Imageamento por Ressonância Magnética Limite: Humans Idioma: En Ano de publicação: 1993 Tipo de documento: Article