The Results of Long-term Follow-up of Total Hip Arthroplasty Using Hydroxyapatite-coated Cups / 대한고관절학회지
Hip & Pelvis
; : 209-215, 2015.
Article
em En
| WPRIM
| ID: wpr-198808
Biblioteca responsável:
WPRO
ABSTRACT
PURPOSE:
The aim of this study was to report the long-term outcome and the failure mechanism of cementless total hip arthroplasty (THA) using hydroxyapatite (HA)-coated acetabular cup. MATERIALS ANDMETHODS:
From January 1992 to May 1994, a total of 123 consecutive cementless primary THAs were performed using a HA-coated acetabular cup with metal-on-polyethylene articulation. We retrospectively evaluated 66 hips available for follow-up at a mean 18.3 years (range, 10.4-23.6 years). The survival analysis was performed by the Kaplan-Meier method. We defined end point as any failure that required a reoperation of acetabular component.RESULTS:
Thirty-nine of 66 hips (59.1%) were defined as a failure for progressive acetabular osteolysis or aseptic loosening of the cup. Acetabular osteolysis was observed in 47 hips (71.2%) and 33 hips (50.0%) were revised because of cup loosening. The Kaplan-Meier method showed the survival rate of the acetabular cup to be 46.3% at 15 years and 34.8% at 20 years for any failure that required a reoperation of acetabular component.CONCLUSION:
The long-term survival rate of THA using HA-coated acetabular cup was unsatisfactory, and it was attributed to vulnerable property of HA coating and progressive osteolysis.Palavras-chave
Texto completo:
1
Base de dados:
WPRIM
Assunto principal:
Osteólise
/
Reoperação
/
Taxa de Sobrevida
/
Estudos Retrospectivos
/
Seguimentos
/
Durapatita
/
Artroplastia de Quadril
/
Quadril
/
Acetábulo
Tipo de estudo:
Observational_studies
/
Prognostic_studies
Idioma:
En
Ano de publicação:
2015
Tipo de documento:
Article