1.
Appl Radiat Isot
; 134: 340-350, 2018 Apr.
Artigo
em Inglês
| MEDLINE
| ID: mdl-29146526
RESUMO
The sum-peak method, an absolute measurement technique that uses coincidence counting and γ-ray spectrometry, was applied to activity standardization of 152Eu sources. The decay branch used was that of 152Sm, as it is almost entirely electron capture and exhibits X-rays with a high probability of coincidence with the 121.8keVgamma rays, as evidenced by the matrix technique used for complex decay schemes, a method for deriving counting rate equations describing coincidence summing of gamma and X-rays. The result was compared with the calibration performed by ionization chamber traceable to BIPM. The results presented uncertainty values of 0.50% (k = 1).