1.
Am J Emerg Med
; 36(3): 467-469, 2018 Mar.
Artigo
em Inglês
| MEDLINE
| ID: mdl-29306645
RESUMO
A patient presented with symptomatic third degree atrioventricular block requiring emergency transvenous pacemaker placement. During the procedure, wireless digital radiographs tracked the position of the pacemaker electrode, which repeatedly missed the target right ventricle. The patient was then rolled to left lateral decubitus position and the electrode was advanced into the right ventricle, achieving electrical capture, hemodynamic stability, and symptom resolution. We review the published literature on transvenous pacemaker placement and identify two innovations: left lateral decubitus position to facilitate catheter placement and wireless digital radiography for procedure guidance.
Assuntos
Marca-Passo Artificial , Implantação de Prótese/métodos , Radiografia Intervencionista , Tecnologia sem Fio , Idoso , Bloqueio Atrioventricular/terapia , Humanos , Masculino