Artif Organs ; 46(1): 71-82, 2022 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-34287976

RESUMO

The objective of this study is to identify the preload and afterload sensitivity of the ReinHeart TAH 2.0. For adequate left-right flow balance, the concept of a reduced right stroke volume (by about 10%) and active adaption of the right diastole duration are evaluated concerning the controllability of the flow balance. This study used an active mock circulation loop to test a wide range of preload and afterload conditions. Preload sensitivity was tested at atrial pressures (APs) between 4 and 20 mm Hg. Left afterload was varied in a range of 60-140 mm Hg mean aortic pressure (MAP), right afterload was simulated between 15 and 40 mm Hg. Four scenarios were developed to verify that the flow difference fully covers the defined target range of 0-1.5 L/min. Although a positive correlation between inlet pressure and flow is identified for the right pump chamber, the left pump chamber already fills completely at an inlet pressure of 8-10 mm Hg. With increasing afterload, both the left and right flow decrease. A positive flow balance (left flow exceeds right flow) is achieved over the full range of tested afterloads. At high APs, the flow difference is limited to a maximum of 0.7 L/min. The controllability of flow balance was successfully evaluated in four scenarios, revealing that a positive flow difference can be achieved over the full range of MAPs. Under physiological test conditions, the linear relationship between flow and heart rate was confirmed, ensuring good controllability of the TAH.


Assuntos
Circulação Sanguínea , Coração Artificial , Desenho de Prótese , Pressão Sanguínea , Frequência Cardíaca , Hidrodinâmica , Modelos Cardiovasculares
2.
Artif Organs ; 45(4): 364-372, 2021 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-33001469

RESUMO

A total artificial heart (TAH) must be designed to autonomously balance the flows of the systemic and pulmonary circulation to prevent potentially lethal lung damage. The flow difference between the systemic and pulmonary circulation is mainly caused by the bronchial (arteries) shunt flow and can change dynamically. The ReinHeart TAH consists of only one actuator that ejects blood alternately from the right and left pump chamber. This design entails a coupling of the right and left stroke and thus, complicates the independent adaptation of the right and left flow. In this experimental study on the ReinHeart TAH, four concepts to keep the flows well balanced were investigated using an active mock circulation loop for data acquisition. Three concepts are based on mechanical design changes (variation of pusher plate shape, flexible right pump chamber housing, and reduced right stroke volume) to achieve a static flow difference. In combination with these static concepts, a concept influencing the ratio of systole and diastole duration to respond to dynamic changes was studied. In total, four measurement series, each with 270 operating points, to investigate the influence of circulatory filling volume, heart rate, bronchial shunt flow, and lung resistance were recorded. In the course of this study, we introduce a concept deviation indicator, providing information about the efficiency of the concepts to balance the flows based on changes in lung's blood pressures. Furthermore, the distribution of the measured data was evaluated based on bubble plot visualizations. The investigated variation of the right pusher plate shape results in high lung pressures which will cause lethal lung damage. In comparison, a flexible right pump chamber housing shows lower lung pressures, but it still has the potential to damage the lungs. Reducing the stroke volume of the right pump chamber results in proper lung pressures. The flow balance can dynamically be influenced with a positive effect on the lung pressures by choosing a suitable systole-diastole-ratio. The results of this study suggest that an adequate right-left flow balance can be achieved by combining the mechanical concept of a reduced right stroke volume with an active control of the systole-diastole-ratio.


Assuntos
Coração Artificial , Modelos Cardiovasculares , Desenho de Prótese , Circulação Pulmonar/fisiologia , Pressão Sanguínea/fisiologia , Artérias Brônquicas/fisiologia , Diástole/fisiologia , Frequência Cardíaca/fisiologia , Humanos , Volume Sistólico/fisiologia , Sístole/fisiologia