1.
Radiat Prot Dosimetry ; 193(3-4): 176-184, 2021 May 17.
Artigo em Inglês | MEDLINE | ID: mdl-33823534

RESUMO

The aim of this study is to characterise the neutron flux generated directly behind targets used in medical cyclotrons. The characterisation process aims at determining the feasibility of using the generated neutrons for research purposes in neutron activation analysis. The study was performed by activating gold foils placed directly behind the cyclotron targets. The thermal and epithermal neutron flux were found to be 4.5E+05 ± 8.78E+04 neutrons cm-2 s-1 and 2.13E+06 ± 8.59E+04 neutrons cm-2 s-1, respectively. The flux value is the same order of magnitude listed in the manual produced by the cyclotron manufacturer. The results are encouraging and show high potential for using the cyclotron facility as a thermal neutron source for research purposes. However, it is important radiation protection procedures be followed to ensure the safety of researchers due to the high gamma dose rate measured directly behind the target at 2.46 Sv/h using an OSL chip during the beam on time.


Assuntos
Terapia por Captura de Nêutron de Boro , Proteção Radiológica , Ciclotrons , Raios gama , Nêutrons